3.4.7 Pipework reducers
Always avoid extreme step changes in pipe size. Use pipework reducers, where there is a large
difference between pipework size and meter flanges.
3.4.8 Flexible connections
Flexible connections can be used but because of the high flow rates associated with large
diameter meters, it is recommended that flexible connections are not used on meters larger
than size 80.
3.4.9 Heating and insulation
Heating
Heating
Heating
Heating
The meter can be heated by electrical tape (or similar) as shown. ONLY heat the meter in the
area marked A.
Insulation
Insulation
Insulation
Insulation
The meter can also be insulated to a maximum depth as shown (B). Do not insulate above this
depth as this will cause the electronics to overheat.