6
TECHNICAL DATA
26
OPTIFLUX 5000
www.krohne.com
03/2017 - 4000686204 - MA OPTIFLUX 5000 SW R04 en
6.2 Technical data
INFORMATION!
•
The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local sales office.
•
Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Downloadcenter).
Measuring system
Measuring principle
Faraday's law
Application range
Electrically conductive fluids
Measured value
Measured value
Measured value
Measured value
Primary measured value
Flow velocity
Secondary measured
value
Volume flow, mass flow, electrical conductivity, coil temperature
Design
Features
Sandwich version with optimized flow tube
Modular construction
The measurement system consists of a flow sensor and a signal converter.
It is available as compact and as separate version. More information about
the signal converter can be found in the documentation of the signal
converter.
Compact version
With IFC 050 converter : OPTIFLUX 5050 C
With IFC 100 converter : OPTIFLUX 5100 C
With IFC 300 converter: OPTIFLUX 5300 C
DN2.5...15 / 1/10...1/2" is only available with aluminum converter housing
Remote version
In wall (W) mount version with IFC 050 converter : OPTIFLUX 5050 W
In wall (W) mount version with IFC 100 converter : OPTIFLUX 5100 W
In field (F), wall (W) or rack (R) mount version with IFC 300 converter :
OPTIFLUX 5300 F, W or R
Nominal diameter
DN2.5...100 / 1/10...4"
Measuring accuracy
Maximum measuring
error
IFC 050: down to 0.5% of the measured value
±
1 mm/s
IFC 100: down to 0.3% of the measured value
±
1 mm/s
IFC 300: down to 0.15% of the measured value
±
1 mm/s
The maximum measuring error depends on the installation conditions.
For detailed information refer to
Measuring accuracy
on page 31.
Repeatability
±
0.1% of MV, minimum 1 mm/s
Calibration
Standard:
Standard:
Standard:
Standard: 2 point calibration by direct volume comparison.
Optional:
Optional:
Optional:
Optional: special calibration on request.