Installation and operating instructions
38 BATCHFLUX
7.2 Test of measuring range Q, Fct. 2.1
Please note!
Do not open the housing of the
BATCHFLUX IFM 5015 K.
Danger of contamination with substances
likely to destroy the moisture barrier of the
electronic equipment (e.g. if CIP or SIP
cleaned from the outside).
Therefore, please contact your KROHNE
Service engineer before you open the
housing.
Switch off power source
before opening the housing
•
For this test a measured value can be
simulated in the range of -110 to +110
percent of Q
100%
(for set full-scale range
see Fct. 1.1 FULL SCALE).
•
Power the system.
•
Press the following keys for this test:
Keys Displayed
Description
→
If “YES“ is set under Fct. 3.4 ENTRY CODE, key-in the
9-keystroke CODE 1:
→
→
→
↵
↵
↵
↑
↑
↑
Fct.
1.0
OPERATION
↑
Fct. 2.0
TEST
→
Fct. 2.1
TEST Q
→
SURE
NO
↑
SURE
YES
↵
0
PERCENT
Current, pulse and status outputs indicate the
corresponding values.
± 10
PERCENT
↑
or
↓
± 50
PERCENT
Select with keys
↑
and
↓
± 100
PERCENT
± 110
PERCENT
↵
Fct. 2.1
TEST Q
End of test; actual measured values again present at
outputs.
(2x) 3x
↵
- - - - - - -
- - - - - / - - -
Measuring mode
7