Installation and operating instructions
24 BATCHFLUX
Fct. Text
Description
and
settings
3.3 ZERO
SET
Zero calibration
Note: Carry out only when flow
is “0“ and measuring tube is
completely filled!
Precautionary query
• CALIB. NO
Press key
↵
,
return to
Fct. 3.3 “ZERO“.
• CALIB. YES
Press key
↵
to start calibration.
Duration approx. 10 seconds.
Current flowrate displayed in
the selected unit (see Fct. 1.4
“DISP.FLOW“)
A “WARNING“ sign appears
when flowrate
“>
0
“.
Confirm with key
↵
.
• STORE NO (do not store
new zero value)
• STORE YES (store
(store new zero value)
Press key
↵
to return to
Fct. 3.3 “ZERO SET “.
3.4 ENTRY
CODE
Entry code required to enter
setting mode?
• NO (= entry with
→
only)
• YES (= entry with
→
and
Code 1:
→
→
→
↵
↵
↵
↑
↑
↑
)
Press key
↵
to return to
Fct. 3.4 “ENTRY CODE“.
3.5 USER
UNIT
Set any required unit for
flowrate and counting
→
TEXT
VOL.
Set text for required
flowrate unit
(max. 5 characters)
Factory-set: “Liter“ (= Liters).
Characters assignable to
each place:
• A-Z, a-z, 0-9, or „ – “
(= blank character)
Press key
↵
to transfer to
subfunction
“FACT. VOL.“.
→
FACT.
VOL.
Set conversion factor (F
M
)
for volume
Factory set: “1.00000 E+3“ for
“Liter“ (exponent notation,
here:10
3
).
Factor F
M
= volume per 1m
3
.
Setting range
• 1.00000 E-9 to 9.99999 E+9
(= 10
-9
to 10
+9
)
Press key
↵
to transfer to
subfunction
“TEXT TIME“.
Fct. Text
Description and settings
→
TEXT
TIME
Set text for required flow
unit
(max. 3 characters)
Factory-set: “hr“ (= hour).
Characters assignable to
each place:
• A-Z, a-z, 0-9, or “ – “
(= blank character)
Press key
↵
to transfer to
subfunction
“FACT.TIME“.
→
FACT.
TIME
Set conversion factor (F
T
)
for time
Factory set: “3.60000 E+3“
for “hour“ (exponent notation,
here: 3.6 x 10
3
).
Factor F
T
: set in seconds.
Setting range
• 1.00000 E-9 to 9.99999 E+9
(= 10
-9
bis 10
+9
)
Press key
↵
to return to
Fct. 3.5 “USER UNIT“.
4