10 Operation

10.3 Decommissioning

10.2.2

Checking the function

1.

 Check the flow direction of the flowmeter.

2.

 Check the installation and installation position of the pick up and of the temperature sensor.

3.

 Check leaks at the pipe thread/connecting flange.

4.

 Check the electrical installation. Observe the operating instructions for pick up and electronic unit.

5.

 Check the power supply. Observe the operating instructions for pick up and electronic unit.

10.2.3

Commissioning the flowmeter

ATTENTION

Increased wear and/or blockade through solid particles or abrasive fine particles in the liquid.

► Protect the flowmeter through correspondingly dimensioned operating filters in the pipe system.

ATTENTION

Measuring error when the pressure drops below the minimum pressure outlet at the outlet.

► Ensure that sufficient counter-pressure exists at the outlet by routing the piping upwards.

ATTENTION

Measuring error through gas inclusion in the pipe system.

► Before commissioning, make sure that the flowmeter is filled.

► Vent the pipe system.

Requirement:

ü

The ambient conditions correspond to the operating data

ü

Flowmeter installed correctly in the pipe system 

Ä

 Installation, removal, Page 23

ü

Flowmeter connection to the pipe system is stress-free

ü

Pipe system is free of impurities

ü

Pipe system vented

ü

Shut-off devices in the feed line and outlet opened

 Switch on the measuring system.

ð

The flowmeter measures when the pick up generates a signal.

10.3 Decommissioning

10.3.1

Switching off the flowmeter

ATTENTION

Damage to the measuring unit.

In the case of strong changes to the flow rate (for example, rapid shutdown, pulsations, etc.) there is a
danger of excessive pressure differences.

► Ensure that the pressure loss of the flowmeter does not exceed the values specified on the rating

plate or the substitute operating data.

 Ensure that the pressure loss of the flowmeter does not exceed the values specified on the rating

plate or the standardized substitute operating data 

Ä

 Technical data, Page 8.

ð

When the flow through the flowmeter is stopped, the generation of the signal stops automatically.
No further measures are required to switch off.

Operating instructions

OIO 08en-GB Edition 2020-01

29

Содержание OMH Series

Страница 1: ...Operating instructions KRAL flowmeters Series OMH OIO 08en GB Edition 2020 01 Original instructions www kral at...

Страница 2: ...emoval 23 8 3 Installing the flowmeter 23 8 3 1 Protecting the flowmeter against soiling 23 8 3 2 Installation types 23 8 3 3 Installation recommendation 24 8 3 4 Installing the flowmeter 25 8 4 Remov...

Страница 3: ...rques for screw plugs with thread meas ured in inches and elastomer seal 45 16 3 Tightening torques and widths for threaded rings 46 16 4 Contents of the Declaration of Conformity 46 Operating instruc...

Страница 4: ...rately at the beginning of the individual chapters in these instructions The following table provides an overview Target group Activity Qualification Transport personnel Transporting unloading setting...

Страница 5: ...according to EU Directive 2014 68 EU o Corresponding operating instructions for accessories 2 Safety 2 1 Proper use o Flowmeter solely for flow measurement of lubricating liquids that are chemically...

Страница 6: ...ions for the electronic unit and the sensors carefully and observe them o Have work only carried out by qualified personnel trained personnel o Wear personal protective equipment and work carefully o...

Страница 7: ...tion D With recognition of flow direction with temperature compensation X Special design 5 Material of bearing A Steel B Hybrid C Stainless steel D All ceramic X Special design 6 Material of seal A FP...

Страница 8: ...pulses Strong pressure pulsations in the system can reduce the service life of the flowmeter 4 3 Maximum values The following table shows the respective maximum values that however may not occur simul...

Страница 9: ...r 12 16 26 34 42 73 K factor P l 7296 2560 1014 302 167 87 6 Millilitre pulse ml P 0 14 0 39 0 99 3 31 5 98 11 42 Pulse frequency o f4 Qmax Hz 1824 1920 2535 2643 2923 4380 o f4 Qrated Hz 1216 1280 16...

Страница 10: ...kg 7 0 12 0 28 0 54 5 80 5 148 0 Tab 4 Dimensions and weights pipe thread connection 4 6 2 DIN flange L2 L1 D L3 L3 TK D DN M18x1 Fig 5 Dimensional drawing DIN flange connection DN Nominal diameter fl...

Страница 11: ...Total length L2 Length of the flowmeter without connections L3 Flange thickness Unit 013 020 032 052 068 DN inch 1 2 3 4 1 1 1 2 2 Class 6000 6000 6000 6000 6000 D mm 100 145 180 220 235 L1 mm 150 185...

Страница 12: ...000 5000 10000 50000 5 10 50 200 500 1000 2000 100 mm2 s 2 3 4 10 9 8 7 6 5 0 1 1 15 14 13 12 11 E D l min 100 50 150 D l min 50 45 40 30 25 20 15 10 5 1 35 3 5 2 5 1 5 0 5 0 1 4 5 5 1 2 3 4 1 A Short...

Страница 13: ...5 10 50 200 500 1000 2000 100 mm2 s 6 9 12 30 27 24 21 18 15 0 3 3 45 42 39 36 33 E D l min 100 50 150 D l min 50 45 40 30 25 20 15 10 5 1 35 10 5 7 5 4 5 1 5 0 3 13 5 15 3 6 9 12 1 A Short term oper...

Страница 14: ...5 10 50 200 500 1000 2000 100 mm2 s 20 30 40 100 90 80 70 60 50 1 10 150 140 130 120 110 E D l min 100 50 150 D l min 50 45 40 30 25 20 15 10 5 1 35 35 25 15 5 1 45 50 10 20 30 40 1 A Short term oper...

Страница 15: ...500 1000 2000 100 mm2 s 70 105 140 350 315 280 245 210 175 3 5 35 525 490 455 420 385 E D l min 100 50 150 D l min 50 45 40 30 25 20 15 10 5 1 35 122 5 87 5 52 5 17 5 3 5 157 5 175 35 70 105 140 1 A S...

Страница 16: ...200 500 1000 2000 100 mm2 s 140 210 280 700 630 560 490 420 350 7 70 1050 980 910 840 770 E D l min 100 50 150 D l min 50 45 40 30 25 20 15 10 5 1 35 245 175 105 35 7 315 350 70 140 210 280 1 A Short...

Страница 17: ...1000 2000 100 mm2 s 400 600 800 2000 1800 1600 1400 1200 1000 20 200 3000 2800 2600 2400 2200 E D l min 100 50 150 D l min 50 45 40 30 25 20 15 10 5 1 35 700 500 300 100 20 900 1000 200 400 600 800 1...

Страница 18: ...ll the rotating parts Thanks to the dis placement principle the flowmeter does not require inlet sections and smoothing sections at the feed line and outlet Depending on the customer requirements the...

Страница 19: ...ow rates These are also documented in the enclosed calibration certificate If highest measuring precision is required it is therefore advisable especially at strongly varying flow rates to take these...

Страница 20: ...ransportation immediately to the manufacturer 3 Store the supplied sensors for the installation 4 Dispose of packaging material in accordance with the locally applicable regulations 6 4 Transporting f...

Страница 21: ...s The flowmeter is additionally packed in hermetically sealing anti corrosion paper ATTENTION Damage to equipment and corrosion if stored improperly and during longer standstills Protect the flowmeter...

Страница 22: ...thing o Protective gloves o Safety boots Aids o Castrol Rustilo DWX 33 or other preservative offering compar able protection 1 Paint or spray the preservative onto all plain and unpainted parts 2 At i...

Страница 23: ...life of the flowmeter In order to protect the flowmeter against soiling the manufacturer generally recommends the installation of an operating filter Personnel qualification o Fitter Personal protecti...

Страница 24: ...ion 3 o Bypass with 3 shut off valves for flange connection o With or without operating filter o Dismantling of the flowmeter without operation interruption 4 o Bypass with 3 shut off valves for pipe...

Страница 25: ...Source for measuring error Gas inclusions in the pipe system 8 3 4 Installing the flowmeter Personnel qualification o Fitter Personal protective equipment o Work clothing o Protective gloves o Safety...

Страница 26: ...pply and secure it against being switched back on Flowmeter cooled down to the ambient temperature 1 Drain the pipe system or divert the pumped liquid via bypass 2 Remove the flowmeter 3 Attach protec...

Страница 27: ...shed and cleaned using the flowmeter ATTENTION Damage to device when the pipe threading is screwed in too far Observe the thread length of the flowmeter Use a standard cutting ring screw connection No...

Страница 28: ...he following safety instructions must be observed strictly o Have all work only carried out by authorized qualified personnel o Ensure that the flowmeter is only operated within the operating limits o...

Страница 29: ...flowmeter is filled Vent the pipe system Requirement The ambient conditions correspond to the operating data Flowmeter installed correctly in the pipe system Installation removal Page 23 Flowmeter co...

Страница 30: ...trongly accelerated wear can occur in the flowmeter Notice In cases of doubt the manufacturer recommends already carrying out the first inspection after twelve weeks operation time 11 2 Maintaining fl...

Страница 31: ...mbient temperature Avoid rapid temperature changes o Pumped liquids can be hot poisonous combustible and caustic o Observe the tightening torques Appendix Page 45 o Observe the operating instructions...

Страница 32: ...ing cover 2 in the pick up hole In the process observe the correct position of the support ring and O ring 3 2 4 5 1 8 7 6 15 A 1 6 Carefully slide the screw set into the measuring housing 1 7 Slide t...

Страница 33: ...17 from the measuring screw small 18 12 2 4 Installing seals and ball bearings OMH 020 OMH 032 Personnel qualification o Fitter Personal protective equipment o Work clothing o Protective gloves o Safe...

Страница 34: ...o Plastic hammer o Extractor Requirement Flowmeter removed from system Completion consisting of socket screws flanges and seals removed Pick up inserts removed 1 On the floating bearing end remove th...

Страница 35: ...pull off the ball bearing 25 using an extractor 2 15 16 3 16 15 12 Remove the O ring 15 and support ring 14 from the pick up hole in the bearing cover 3 12 2 6 Installing seals and ball bearings OMH 0...

Страница 36: ...nsure the correct positioning of the wedge lock washer wedge on wedge 11 Apply Loctite 242 to the socket screw 6 and install with mounted washers on the measuring screw small 26 12 Pull the wedge lock...

Страница 37: ...9 Remove the circlip 16 and support ring 15 from the measuring screw large 13 10 Remove the circlip 17 and support ring 18 from the measuring screw small 20 11 Use the extractor to pull the ball bear...

Страница 38: ...ing unit into the measuring hous ing 1 tighten the socket screws with torque 13 Disposal 13 1 Dangers during disposal The following safety instructions must be observed strictly o Neutralize residues...

Страница 39: ...t Use the commissioning filter 3 5 Pick up not connected correctly Check the power supply for the pick up In the process observe the operating instruc tions of the pick up 3 5 Pick up defective Check...

Страница 40: ...ons of the elec tronic equipment 3 5 Filling amount too low Use a flowmeter with a suitable size Use a flowmeter from a suitable series 5 Strongly deviating operating data Use a flowmeter from a suita...

Страница 41: ...9 5 O ring 107 1 Pipe thread connection 817 1 Deep groove ball bearing 107 2 Pipe thread connection 817 2 Deep groove ball bearing 128 Measuring housing 817 3 Deep groove ball bearing 259 Pole wheel 8...

Страница 42: ...earing 080 1 Bearing cover 817 3 Deep groove ball bearing 107 1 Pipe thread connection 817 4 Deep groove ball bearing 107 2 Pipe thread connection 869 2 Circlip 128 Measuring housing 869 4 Circlip 259...

Страница 43: ...bearing 064 2 Supporting ring 818 2 Angular contact ball bearing 080 1 Bearing cover 868 2 Support ring 080 2 Bearing cover 868 4 Support ring 107 1 Pipe thread connection 869 2 Circlip 107 2 Pipe th...

Страница 44: ...21 1 Four point contact bearings 064 2 Supporting ring 821 2 Four point contact bearings 080 1 Bearing cover 868 1 Support ring 080 2 Bearing cover 868 2 Support ring 107 1 Pipe thread connection 868...

Страница 45: ...80 8 8 M 3 0 6 1 5 1 2 1 M 4 1 4 3 0 4 1 2 3 2 M 5 2 7 6 0 8 0 4 8 3 5 4 7 5 M 6 4 7 10 3 14 0 7 6 6 0 8 0 9 M 8 11 3 25 0 34 0 18 4 16 0 22 0 14 M 10 23 0 47 0 68 0 36 8 32 0 43 0 36 M 12 39 0 84 0 1...

Страница 46: ...of the Directive 2006 42 EC The original of the EC Declaration of Conformity is enclosed with the machinery at delivery The machinery fulfils all the relevant provisions of the following directives N...

Страница 47: ...Notes Operating instructions OIO 08en GB Edition 2020 01 47...

Страница 48: ...KRAL GmbH 6890 Lustenau Austria Tel 43 5577 86644 0 E Mail kral kral at www kral at...

Отзывы: