8-7
[0C/0D Transfer elevator motor failure (upper/lower limit position)]
Procedure Points to be checked
Remedy
Checking method
1
Does the transfer
elevator motor operate
at the proper timing
during the power-on
initialization process?
If YES, go to 5.
If NO, go to 2.
2
Is the fuse 5 blown?
If NO, go to 3.
If YES (fuse is
blown), replace the
KMPU board.
Measure the resistance across the ends of
F5 (fuse 5) of the KMPU board with a
digital multimeter.
The fuse is not blown if the resistance is
approximately 0
Ω
.
3
Is the KMPU board
normal?
If YES, go to 4.
If NO, replace the
KMPU board.
Turn the POWER switch OFF.
Connect the jack of another normal
transfer elevator motor to the connector
ELEV-MTR.
Turn the POWER switch ON. If normal,
the transfer elevator motor operates at
the proper timing during the power-on
initialization process.
Turn the POWER switch OFF. Connect
the original transfer elevator motor back
to the connector ELEV-MTR .
4
Is the transport unit set
properly?
If YES, go to 5.
If NO, set the
transport unit
correctly.
5
Is the transfer elevator
motor normal?
If YES, go to 6.
If NO, replace the
transfer elevator
motor.
Turn the POWER switch ON. If normal,
the transfer elevator motor operates at
the proper timing during the power-on
initialization process.
6
Is the transfer elevator
motor overloaded?
If NO, go to 7.
If YES, replace the
KMPU board.
Check as follows
:
Do the transport unit and the
transfer/separation section move up and
down when the transfer elevator motor is
operated? (Normal if YES.)
If YES, go to 7.
If NO, check the next point.
Is an unusual noise emitted during the
above process? (Normal if NO.)
If YES, go to 7.
If NO, check the next point.
When power-on initialization is
performed after pulling out the transport
unit, does the motor shaft of the transfer
elevator motor turn?
If YES, check the load on the transport
unit.
If NO, replace the KMPU board.
7
Is the transfer unit
upper limit sensor
normal?
If YES, go to 8.
If NO, replace the
upper sensor.
Check the following after removing the
transfer elevator motor along with the
transfer unit upper limit sensor and the
bracket.
-To be continued-
Содержание KP3828 Series
Страница 4: ...Chapter 1 Overview ...
Страница 5: ...1 1 1 1 External View External View ...
Страница 17: ...Chapter 2 Principles ...
Страница 40: ...Chapter 3 Installation ...
Страница 47: ...3 7 Step 13 Remove the nut from the drum shaft Fig 3 7 Nut ...
Страница 59: ...3 19 Fig 3 19 AC power terminal box AC cable Breaker Turn on AC power terminal box KMPU board ...
Страница 64: ...Chapter4 Electrical System 4 1 Diagram 4 1 4 2 Electrical Configuration 4 2 4 3 Input Output Signals 4 5 ...
Страница 65: ...Chapter 4 Electrical System ...
Страница 70: ...4 5 4 3 Input Output Signals I O signal block diagram of the KMPU Fig 4 5 KMPU I O Signal Block Diagram ...
Страница 72: ...Chapter 5 Adjustments ...
Страница 88: ...5 16 Fig 5 16 Fig 5 17 ...
Страница 90: ...Chapter 6 Periodic Maintenance ...
Страница 96: ...6 6 6 3 3 Developing Unit Hopper Area Fig 6 3 6 3 4 Cleaning Unit Area Fig 6 4 Cleaning unit ...
Страница 105: ...6 15 2 Wipe the drum surface with the special cloth Fig 6 14 Drum Special cloth ...
Страница 113: ...Chapter 7 Parts Replacement ...
Страница 143: ...7 30 5 Remove the screw from the front bracket and the rear bracket Then remove the lamp Fig 7 48 Screw ...
Страница 145: ...7 32 3 Open the cover remove the screws securing the fuse set and remove the fuse Fig 7 51 Screws Fuse Set ...
Страница 154: ...7 41 7 4 2 DCMTR DRV Board 1 Remove the connectors 2 Remove the screws that secure the DCMTR DRV board Fig 7 64 Screws ...
Страница 172: ...Chapter 8 Troubleshooting ...
Страница 230: ...Appendix I KMPU Circuit Diagram ...