31
4 Other functions
4.2 Small pupil photography
This section describes how to photograph patients with a small pupil diameter.
While this method allows you to photograph patients with a small pupil diameter, there are such disadvan-
tages as smaller field angles or easier occurrence of a flare.
Press the “S.P. button” on the right side panel to change the pupil diameter that can be photographed.
The pupil diameter that can be photographed under each photography mode is shown below.
Photography mode
Standard
S.P.
S.S.P
Mydriatic color
Φ5.5 mm
Φ4.0 mm
Φ3.5 mm
Red Free
Φ4.0 mm
×
×
FA
Φ5.5 mm
Φ4.0 mm
Φ3.5 mm
Non-mydriatic color
Φ4.0 mm
Φ3.5 mm
×
FAF (Non-mydriatic monitoring)
Φ4.0 mm
Φ3.5 mm
×
FAF (Mydriatic monitoring)
Φ5.5 mm
Φ4.0 mm
Φ3.5 mm
¿
pupil of the diameter shown in the table or larger may be photographed.
Two levels of the S.P. mode is available for mydriatic color photography, fluorescence angiography, and
mydriatic monitoring in FAF.
The S.P. button on the right side panel is indicated as shown below:
Normal pupil diameter : Off
S.P.
: Lit
S.S.P.
: Flashing