18 EMC Guidelines
Below cables information are provided for EMC reference.
Cable
Max. cable length,
Shielded/unshielded
Number
Cable
classification
AC Power Line
1.5m
shielded
1 Set
AC Power
Important information regarding Electro Magnetic Compatibility
(EMC)
Oxygen Concentrator needs special precautions regarding EMC and put into
service according to the EMC information provided in the user manual;
Oxygen Concentrator conforms to this IEC 60601-1-2:2014 standard for both
immunity and emissions. Nevertheless, special precautions need to be
observed:
Oxygen Concentrator with no Essential Performance/Following Essential
Performance is intended used in Professional healthcare facility environment.
Essential Performance:
Noise: KSOC-3 Series ≤53dB
;
KSOC-5/KSOC-8 Series ≤54dB;
O
2
density (V/V):93%±3%
WARNING: Portable RF communications equipment (including
peripherals such as antenna cables and external antennas) should be
used no closer than 30 cm (12 inches) to any part of the Oxygen
Concentrator, including cables specified by the manufacturer.
Otherwise, degradation of the performance of this equipment could
result.”
The use of accessories, transmitters and cables other than those specified by
Konsung, with the exception of accessories and cables sold by Konsung of
Oxygen Concentrator as replacement parts for internal components, may
result in increased emissions or decreased immunity of the Oxygen
Concentrator.
WARNING: Use of this equipment Oxygen Concentrator adjacent to
or stacked with other equipment should be avoided because it could
result in improper operation.”