11
7 Models and Performance
“N” Model is feature configuration, which is added air atomization function
on the basic functions.
“T” Model is feature configuration, which is added the function of Remote
Control.
“M” Model is feature configuration, which is added the functions of air
atomization and Remote Control.
All models of this series are equipped with function: current run time, total
running time function, timing, and power off alarm.
Class of safety type: It belongs to CLASS II equipment
Type BF Applied part for safety type
Oxygen concentrator is non AP/APG device
Protection class of case: IPX0
Running mode: continuous operation
Outlet pressure: 0.04MPa
~
0.07MPa
Pressure range of atomization outlet: 60kPa~250kPa
Safety valve pressure range: 15kPa
~
40kPa
Safety valve pressure range: 250kPa
~
280kPa
Overloading protector specification:
KSOC-3/5: 250V AC, 3A
110V AC, 5A
KSOC-8: 250V AC, 5A
110V AC, 8A
8 Working Conditions
Power: AC 220-240V 50/60HZ or AC110V 60HZ
Environment temperature: 5℃
~
40℃
Relative humidity: ≤75%
Air pressure: 86kPa
~
106kPa