Konsung KSOC-8 Скачать руководство пользователя страница 32

 

28 

9

 

EMC 

This product has passed the electromagnetic compatibility test conducted by 

testing center. Please install and use this product according to the following 

electromagnetic compatibility information. Portable and mobile radio 

frequency communication equipment may affect the medical electrical 

equipment. Do not use this product near the places with strong 

electromagnetic interference sources, such as walkie-talkies, CT, MRI and other 

equipment. 

The oxygen concentration of concentrator is 93%±3% when conducting EMC 

test. 

This product uses the ordinary power line RVV 2x0.75, its maximum length is 2 

meters. 

 

WARNING 

 

Use of cables and transducers other than those specified, with the 

exception of cables and transducers sold by the manufacturer of the 

medical electrical equipment as replacement parts for internal 

components, may result in increased emissions of decreased immunity 

of the Oxygen Concentrator. 

 

The Oxygen Concentrator should not be used adjacent to or stacked with 

other equipment. If adjacent or stacked use is unavoidable, the device 

should be observed to verify normal operation. 

 

9.1

 

Electromagnetic emissions 

The medical oxygen concentrator is intended for use in the electromagnetic 

environment specified below. The customer or the user of the concentrator 

should assure that it is used in such an environment. As following: 

Emission test 

Compliance 

Electromagnetic environment 

RF emissions 

CISPR 11 

Group 1 

The medical oxygen concentrator uses RF 

energy only for its internal functions. 

Therefore, its RF emissions are very low 

and are not likely to cause any 

interference in nearby electronic 

equipment. 

RF emissions   

CISPR 11 

Class B 

The medical oxygen concentrator is 

suitable for use in all establishments, 

including domestic establishments and 

those directly connected to the public 

Harmonic emissions   

IEC 61000-3-2 

Class A 

Содержание KSOC-8

Страница 1: ...I Read this manual carefully before use edical Oxygen Concentrator User s Manual...

Страница 2: ...implicitly any right or license to use any of the intellectual properties of us Our company holds the rights to modify update and ultimately explain this manual All illustrations in this manual serve...

Страница 3: ...3 4 Parts and function introduction 9 3 5 Power ON OFF 12 3 6 Oxygen uptake operation 13 3 7 Atomization operation of N M type oxygen concentrator not administered 14 3 8 Other operations 15 3 8 1 Ti...

Страница 4: ...cations 25 7 3 Concentrator models and Technical parameters 25 8 Warranty and Service 27 9 EMC 28 9 1 Electromagnetic emissions 28 9 2 Electromagnetic immunity 29 9 3 Electromagnetic immunity 30 9 4 R...

Страница 5: ...function after heavy physical or mental consumption Concentrator s safety life is five years Contraindications Oxygen poisoning oxygen allergy patients are prohibited 1 1 Application Scope Oxygen conc...

Страница 6: ...2 The following figure shows the process The following chart shows the electrical control process...

Страница 7: ...rohibited Caution No smoking Keep dry Fragile handle with care This side up Increase or decrease Knob ON power Class II equipment OFF power Alternating current Type BF applied part Date of Manufacture...

Страница 8: ...e Unplug the plug cord from the power outlet if long time no use of concentrator Do not pull out the power cord when removing the plug 2 1 Safety tips for oxygen concentrator WARNING Electrical shock...

Страница 9: ...near high frequency disturbing equipment such as speaker MRI or CT etc Do not use parts accessories or adapters that are not approved by the manufacturer Using other humidifier or other accessories n...

Страница 10: ...different power outlet Please pay attention to electricity safety Do not turn on the concentrator if the plug or power lines damaged Ensure to cut off the power when cleaning the concentrator or clea...

Страница 11: ...our company The atomization device and the nasal cannula as accessories provided by our company is the test sample The size of the atomization cup is 6 12mL and the length of the nasal cannula is 1 6...

Страница 12: ...or The unit should be stored in the environment without corrosive air and with good ventilation Be careful to transport or convey don t let it reversal or thwart tilt angle not greater than 5 NOTE Whe...

Страница 13: ...introduction Front view 1 Indicator LED power failure alarm indicator Accidental power off happened while operating the concentrator Humidifier bottle This port is connected to oxygen outlet by a con...

Страница 14: ...lights up in yellow When the temperature of the system exceeds the maximum allowable temperature limit 75 C the oxygen generator stops working The indicator lights yellow and the screen displays the c...

Страница 15: ...reen when using buttons of remote controller 2 Timing 1 hour button 3 Start Stop oxygen button 4 Timing 2 hours button 5 Timing 3 hours button Filter cover Remove this cover you can find the filter Re...

Страница 16: ...lion components Automatic control and fault alarm according to preset working procedures Producing gas treatment system ABS resin Polypropylene Collecting filtering regulating and humidifying the oxyg...

Страница 17: ...IN mark of the bottle and then cover the top part 3 Install the humidification bottle on the humidifier bottle holder of concentrator and secure it with band 4 Press the power switch to position I ON...

Страница 18: ...ied 3 7 Atomization operation of N M type oxygen concentrator not administered 1 Press the power switch to position I ON to turn on the concentrator 2 Connect the atomized mask or mouthpiece to the me...

Страница 19: ...oxygen concentrator LCD will shows the remaining time When remaining time becomes 0 the oxygen concentrator will go to standby status 3 8 2 Remote control operation The function of buttons on the rem...

Страница 20: ...KSOC 8 series oxygen concentrator is 8L min The recommended maximum concentration is 90 96 The recommended maximum flow rate of the KSOC 10 series oxygen concentrator is 10L min The recommended maximu...

Страница 21: ...tion Beep continuous sounding Flashing frequency of the red alarm lamp 1 4Hz 2 8Hz High Please turn off the power immediately If an alarm still exists after confirming that the power supply and connec...

Страница 22: ...e plug to the socket correctly 2 Check the power supply 3 Replace the fuse After turned on the switch the light works but the oxygen concentrator doesn t work 1 Air compressor protection 2 Inlet or ou...

Страница 23: ...ce battery 4 3 Fault code The description of fault code displayed on concentrator is as follows Code Fault Description E01 The pressure cannot reach the required for 25s or longer after startup E02 Du...

Страница 24: ...c shock Person without training or authorization of manufacturer cannot open the cabinet 5 1 Care and Cleaning of cabinet It s helpful for normal use to keep the oxygen concentration clean for long Th...

Страница 25: ...Please clean or replace it in time It is advice that you should clean or replace the filter after each 100 hours WARNING Do not operate oxygen concentrator without filter installation Otherwise it wi...

Страница 26: ...before it can be installed on the filter Recommended cleaning agents are household neutral detergent or vinegar solution 1 V vinegar with 10 V water 5 4 Overload protection When you suspect or determ...

Страница 27: ...aste disposable nasal cannula filter atomization assembly and oxygen concentrator should follow local laws and regulations in case of environmental pollution Reset button Remove the cover by pressing...

Страница 28: ...ed Check the accessories and their packages for any sign of damage Do not use them if any damage is detected The accessory material that contacts the user or other personnel has undertaken the bio com...

Страница 29: ...ent Temperature 5 C 40 C Relative humidity 75 non condensing Barometric pressure 86kPa 106kPa Storage and transportation environment Temperature 40 C 55 C Relative humidity 93 non condensing Barometri...

Страница 30: ...delivered by oxygen concentrator Altitude 800m 800 1500m 1500 3000m 3000 4000m KSOC 8 series 90 82 75 65 KSOC 10 series 90 80 72 63 Impact of outlet pressure on output flow When the outlet nominal pr...

Страница 31: ...ter sold More than 12 month the company provides the parts to repair with reasonable charge if it could not be used The following conditions are not covered by the warranty Damage or deformation of th...

Страница 32: ...ents may result in increased emissions of decreased immunity of the Oxygen Concentrator The Oxygen Concentrator should not be used adjacent to or stacked with other equipment If adjacent or stacked us...

Страница 33: ...d be at least 30 Electrical fast transient burst IEC 61000 4 4 2kV for power supply lines 1 kV for input output lines 2kV for power supply lines Not applied Mains power quality should be that of a typ...

Страница 34: ...lectromagnetic environment specified below The customer or the user of the concentrator should assure that it is used in such an environment As following Immunity test IEC60601 Test level Compliance l...

Страница 35: ...dio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey sho...

Страница 36: ...38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equatio...

Страница 37: ......

Отзывы: