19
Atomization
amount is
insufficient or
cannot be
atomized while
using the
atomization
1.
The atomizing
assembly installed
improperly or
incorrectly;
2.
The atomizing
assembly outlet is
blocked;
1.
Install the assembly properly and
correctly
2.
Replace the atomizing assembly
If this condition still exits, stop using
the atomization assembly and
contact the supplier immediately.
Remote controller
cannot work.
1.
The concentrator is
out of the remote
range.
2.
Lower battery
capacity or no
power.
1.
Using remote controller within
the remote range.
2.
Replace battery
4.3
Fault code
The description of fault code displayed on concentrator is as follows:
Code Fault Description
E01 The pressure cannot reach the required for 25s or longer after startup.
E02 During operation, the pressure suddenly drops below the limit pressure
(20kPa)
E05 The pressure exceeds the limit pressure (260kPa) during operation.
E31 Cannot receive data from oxygen sensor
E35 Compressor temperature control switch is not connected or exceeds the
allowable range
If it’s not in the above cases and there is still no oxygen output, please
contact the distributor or the manufacturer.
Non-professional maintenance personnel or personnel without our
authorization are strictly prohibited to open the concentrator cabinet for
maintenance.