[ViZion DR+ 1417V2 Wireless Panel Calibration Manual]
[Ultra 4.2.9.0]
11
Customer Support
1.800.366.5343
–
2.2 iDetector Interface
After clicking
Utility
under the panel’s Device
settings, the iDetector interface window will
open.
2.2.1 Connecting into the Panel from
iDetector
1. Click to highlight
DEFAULT
under the
name column and then click
Connect
.
Warning
: before exiting iDetector,
DO
NOT CLICK THE RED X
in the top
right corner, without first going to
Home
tab, click
Close
, then click
the
X
and select
Override
Config
.
2. The Acquire tab is selected by default
when you first connect to the panel in
iDetector. Click on the
Detector
tab.
2.2.2 Panel Calibration Setup
1. On the
Acquire
tab in iDetector,
uncheck the
Defect
,
Gain
, and
Offset
checkboxes
in that order
.
Note
: during the calibration, please do not
move the detector once calibration
has been started and keep Link
cable connected to the detector.
2. Select the
Calibrate
tab, check the
radio button for
PostOffset
, and click
Start Generate Templates
.