©2009 KONICA MINOLTA SENSING,INC.
9242-4895-10 AJFAPK 1
Specifications and appearance shown herein subject to change without notice.
•
Company names and product names used herein are trademarks or registered trademarks of their respective com-
•
panies.
Screens shown are for illustration purpose only.
•
Specifications of KONICA MINOLTA RANGE5
Measuring method
Triangulation by light sectioning method
Light source
Semiconductor laser, Wavelength: 660 nm
Laser class
Class 2 (IEC 60825-1 Edition 2)
Number of pixels taken
1.31 megapixels (1280 x 1024)
Measurement distance
450 to 800 mm
Measure-
ment
range
(mm)
Distance
450 mm
800 mm
Direction
X × Y
150×188
267×334
Z
109
194
XY direction
measurement interval
0.16
0.28
Accuracy (Distance be-
tween balls)*
1
±80 μm
Precision (Z, σ)*
2
8 μm
Auto-focus
Provided
Auto-exposure
Provided
Scan time
Approx. 2 sec. or more (1 scan)
Preview function
Approx. 0.4 sec./scan
Scanning ambient
lighting condition
500 lx or less
Output interface
USB 2.0 High Speed
Power
Included AC Adapter
Input voltage: 100 to 240 VAC (50/60 Hz), Rat-
ing: 1.4 A (at 100 VAC input)
Dimensions
295 (W) x 190 (H) x 200 (D) mm (Excluding grips and lens)
Weight
Approx. 6.7 kg
Operating temperature/
humidity range
10 to 40°C; Relative humidity 65% or less (no condensation)
Storage temperature/
humidity range
-10 to 50°C; Relative humidity 85% or less
(at 35°C, no condensation)
*1 When measuring the distance between balls for a ball bar (two balls) as defined in VDI/VDE 2634-2 under Konica
Minolta Sensing measurement conditions.
Konica Minolta Sensing measurement conditions: Temperature: 20±1°C, Measurement distance: 450 mm; Warm-
up: 20 min.; Software used: Konica Minolta processing software; After calibration of the instrument; Measurement
target: Konica Minolta standard target (2 balls); Arrangement of measurement target: Konica Minolta standard ar-
rangement (10 positions inside measurement space); Uncertainty of standard target not included.
*2 Measurement conditions: Temperature: 20±1°C, Measurement distance: 450 mm; Warm-up: 20 min.; Measurement
target: Konica Minolta standard planar chart; Software used: Konica Minolta processing software; 1 σ value
*3 Using Konica Minolta processing software
Specifications of 3D Data Processing Software RANGE VIEWER
<Major Functions>
Readable data
formats
Readable data formats
Konica Minolta proprietary formats: .rgv (Single-scan data group), .rvm
(more than one piece of data)
Data output
STL (binary), ASCII point group
Konica Minolta proprietary formats: .rgv, .rvm, .rmk
Measurement
functions
Monitor image, Preview, AF/AE, Measurement
Control of rotating stage (Konica Minolta optional accessory)
E di t ing f unc-
tions
Data alignment, Data merging (integration), Point group deletion
Drawing
Point group shading
System configuration
KONICA MINOLTA
RANGE5
AC Adapter
USB Cable
(5m)
RA-A101
: Europe only (3m)
Marker Sheet Trial Pack
(500 markers )
RANGE VIEWER
Personal
Computer
(commercial product)
Standard accessories
Optional accessories
3D Data Processing Software
Rotating Stage Kit (3 kg)
Rotating Stage Kit (20 kg)
Platform Set for Tripod
RA-A151
Measuring Stand
Platform Set for
Measuring Stand
RA-A201
Marker Sheet W
RA-A206
(2500 markers)
Carrying Case
RA-A213
Reverse engineering/
testing software
Calibration Sheet
Calibration Chart
RA-A105
RA-A107
Calibration Base
RA-A106
Dolly
Tripod
<Operating Environment>
OS
Windows® Vista Business SP1(64bit),
Windows® XP Professional x64 Edition SP2 (64bit)
CPU
Core2Duo, Xeon or higher
RAM
4 GB or more
Display
Graphics display capability of 1280 x 1024 or more
Graphics board
OpenGL compatible board (Konica Minolta verified compatible board recom-
mended)
Interface
USB 2.0 port
SAFETY PRECAUTIONS
Read all safety and operating instructions before operating
the KONICA MINOLTA RANGE5.
Use only a power source of the specified rating.
Improper connection may cause a fire or electric shock.
Do not stare into the laser beam.
(MAX. 34mW 660nm , CLASS 2 (IEC) LASER PRODUCT)
34mW
http://www.konicaminolta.com/instruments/about/network
Addresses and telephone/fax numbers are subject to change without notice. For the latest contact information,
please refer to the KONICA MINOLTA SENSING Worldwide Offices web page .
Registration No. : YKA 0937154
Registration date : March 3, 1995
Registration No. : JQA-E-80027
Registration date : March 12, 1997
Osaka, Japan
EMail : [email protected]
Web : http://konicaminolta.com/instruments
Konica Minolta Sensing Americas, Inc.
New Jersey , U.S.A.
Phone:
888-473-2656 (in USA), 201-236-4300 (outside USA)
Fax:
201-785-2482
EMail : [email protected]
Web : http://se.konicaminolta.us/3d
Konica Minolta Sensing Europe B.V.
German Office
München, Germany
Phone:
+49(0)89 4357 156 0
Fax:
+49(0)89 4357 156 99
EMail : [email protected]
Web : http://www.konicaminolta.eu
French Office
Roissy CDG, France
Phone :
+33(0)1 493-82519
Fax:
+33(0)1 493-84771
EMail : [email protected]
Web : http://www.konicaminolta.eu
UK Office
Milton Keynes, United Kingdom
Phone :
+44(0)1908 540-622
Fax:
+44(0)1908 540-629
EMail : [email protected]
Web : http://www.konicaminolta.eu
Italian Office
Milan, Italy
Phone :
+39 02 39011.425
Fax:
+39 02 39011.223
EMail : [email protected]
Web : http://www.konicaminolta.eu
Polish Office
Wroclaw, Poland
Phone :
++48(0)71 33050-01
Fax:
+48(0)71 734 52 10
EMail : [email protected]
Web : http://www.konicaminolta.eu
Konica Minolta (CHINA) Investment Ltd.
SE Sales Division
Shanghai, China
Phone:
+86-021-5489 0202
Fax:
+86-021-5489 0005
Konica Minolta Sensing Singapore Pte Ltd.
Singapore
Phone:
+65 6563-5533
Fax:
+65 6560-9721