E 5
English
Error
Messages
Cause
Solution
Readout
and "A"
alternately
flashing
The measurement
results are unreliable
due to motion of the
patient during mea-
surement. The stated
accuracy of measure-
ment is invalid.
Properly insert a fingertip
1
into the finger holder. Avoid
moving the finger in question.
Warm the fingertip and/or
2
shut out surrounding light or
electromagnetic noise.
Flashing
The pulse rate exceeds
the upper limit (230 bpm)
of the measurement
range.
Measurement cannot be per-
formed when the upper limit
of the measurement range is
exceeded.
Flashing
The pulse rate is below
the lower limit (30 bpm)
of the measurement
range.
Measurement cannot be per-
fomed when the lower limit of
the measurement range is not
reached.
Oxygen
saturation
readout
flashing
The oxygen saturation
(SpO
2
) is below the
lower limit (70%) of the
guaranteed accuracy
range.
The stated accuracy
of oxygen saturation
measurements is
invalid.
This message indicates that the
oxygen saturation is out of the
range in which the stated ac-
curacy of measurements is valid.