Measuring geometry

Detector
Wavelength range
Wavelength pitch
Half bandwidth
Photometric range
Light source
Measurement time
Illumination/
measurement areas

Repeatability

Inter-instrument
agreement

Temperature drift

UV adjustment
Specimen conditions
for transmittance
measurements
Interface
Power
Operation temperature/
humidity range (*1)
Storage temperature/
humidity range
Dimensions (W

×

H

×

D)

Weight
Standard accessories

Optional accessories

*1 Operating temperature/humidity range of products for North America : 13 to 33 C, relative

humidity 80% or less (at 31 C) with no condensation

Specifications are subject to change without notice.

55

12

7

271

500

128.5

143

259

247

482

41.8

Dimensions

(Units:mm)

Reflectance: di:8 , de:8 (diffuse illumination/8 viewing angle);
SCI (specular component included )/SCE(specular component
excluded) switchable; meets CIE, ISO, ASTM, and DIN standards.
Transmittance: di:0 , de:0 (diffuse illumination/0 viewing angle)
Silicon photodiode array with flat holographic grating
360 to 740nm
10nm
Approx. 14nm average
0 to 200%; Resolution: 0.001%
Pulsed xenon arc lamp

0.6 to 0.8 sec. (to start of data output)

Reflectance: Changeable between LAV, MAV, and SAV

LAV: ø28mm illumination/ø25.4mm measurement
MAV: ø11mm illumination/ø8mm measurement
SAV: 5

×

7mm illumination/3

×

5mm measurement

Transmittance: Approx. ø20mm
When white calibration plate is measured 30 times at 10-sec.
intervals after white calibration has been performed:
Spectral reflectance: Standard deviation within 0.05%
Chromaticity: Standard deviation within

Δ

E*ab0.005

When black tile(BCRA Series II ; reflectance: 1%)is measured 30 times
at 10-second intervals after white calibration has been performed:
Spectral reflectance: 380 to 740nm: Standard deviation within 0.02%

360 and 370nm: Standard deviation within 0.04%

Chromaticity: Standard deviation within

Δ

E*ab0.05

mean

Δ

E*ab0.08 (typical) Average for 12 BCRA Series II color tiles.

Max

Δ

E*ab0.3(corresponds to approx.

Δ

E*

CMC

0.2) for any of 12 BCRA

Series II color tiles compared to values measured with master body ; LAV.
Spectral reflectance: Within

±

0.10%/ C,

Color difference: Within

±Δ

E*ab0.05/ C

Computer controlled: continuously variable
Sheet, plate, or liquid form up to a maximum thickness of
approximately 50mm

RS-232C standard; baud rates: 1200,2400,4800,or 9600
AC 100V/120V/230V 50/60Hz (using included AC adapter)
13 to 33 C, relative humidity 80% or less with no condensation

0 to 40 C, relative humidity 80% or less with no condensation

271

×

259

×

500mm(10-11/16

×

10-3/16

×

19-11/16 in.)

18kg(39.7 lb.)
White Calibration Plate; Target Mask(3

×

5mm);

Target Mask(ø8mm); Target Mask(ø25.4mm);
Zero Calibration Box; AC Adapter;
RS-232C Cable(9-pin, 2m);
Accessory Case CM-A66
Transmittance Specimen Holder; Glass Cells(2mm, 10mm, 20mm);
Plastic Cells(2mm, 10mm, 20mm);
Transmittance Zero Calibration Plate; Color Tiles;
RS-232C Cable(9-pin, 5m)

Optional Accessories

Transmittance Specimen
Holder CM-A96

Holds specimens in place for trans-
mittance measurements. Maximum
specimen thickness: 22.5mm

Glass Cell
CM-A97/CM-A98/CM-A99

Hold liquid specimens for transmit-
tance measurements. Optical path
lengths: 2mm (CM-A97), 10mm
(CM-A98), and 20mm (CM-A99)

Plastic Cells CM-A130(2mm),
CM-A131(10mm),CM-A132(20mm)
are also available.

Transmittance Zero
Calibration Plate CM-A100

For performing zero calibration for
transmittance measurements.

Color Tiles

14 color tiles are available: White, Pale grey,
Middle grey, Difference grey, Deep grey, Deep
pink, Red, Orange, Bright yellow, Green,
Difference green, Cyan, Deep blue, Black. Original
materials of these tiles are supplied by BCRA.

Specifications (CM-3700d)

SAFETY PRECAUTIONS

For correct use and for your safety, be sure to read the instruction

manual before using the instrument.

Always connect the instrument to the specified power supply

voltage. Improper connection may cause a fire or electric shock.