Chapter 2 Before the Repair
2-31
2.7
Handling the Cassette
Please be familiar with the following cautions when handling the REGIUS cassettes (RC-110 series).
Maintaining the image quality
•
Be cautious no scratch or foreign object goes on the read surface of the plate.
•
If you find any foreign object on the plate read surface, immediately clean it.
•
Use a glove when handling a plate, just as handling the film.
•
Do not expose the plate to the sun light.
•
When you work on the plate under a fluorescent light in situation like cleaning, do it as fast as
possible, and return it to the cassette as soon as it is completed.
•
Do not scratch or smudge the front plate.
Transportation/storage
•
Please follow following specification for temperature and humidity.
•
Temperature: Before opening the plate: 10 ~ 40°C
After opening the plate: 10 ~ 30°C
•
Humidity:
80% RH or less (no condensation)
•
Do not expose to radiation such as X-ray,
γ
-ray, etc.
•
Place it so the cassette will be horizontal when transporting or storing in the original box. The back
plate may distort when the cassette is placed vertically for long period or in high temperature.
•
Always tore the opened plate in the cassette.
•
Do not store with back plate and front plate separated. Especially, the back plate might warp when it
is stored diagonally by itself.
•
Do not drop or bump the cassette.
•
Store the operational cassettes vertically in the rack as conventional cassettes.
Confirmation before inserting
Confirm following before inserting the cassette to prevent the trouble of cassette transportation action in
the REGIUS MODEL 110.
•
Confirm that the lock mechanism is normal, and there is no distortion or breakage. Do not insert the
cassette with abnormality into the REGIUS MODEL 110.
•
Confirm that there is no foreign object, such as a paper, wedged on the memo clip on the back
plate.
•
Confirm that there is no foreign object, such as clips, between the front plate and the back plate.
•
Confirm that there is no tape affixed on the back plate to ensure good absorption of the back plate to
the absorption plate.
•
Confirm that barcode is not dirty.
•
Confirm that it is locked.
Disposal
•
Request a disposal to an authorized industrial waste disposal service when disposing.
Follow the local regulations/rules regarding the disposal methods.
Содержание 0902
Страница 2: ......
Страница 6: ......
Страница 12: ......
Страница 30: ...2 10...
Страница 54: ......
Страница 84: ...Chapter 3 Troubleshooting 3 30 3 2 19 Confirm the H sync Signal To be stated...
Страница 85: ...Chapter 3 Troubleshooting 3 31 3 2 20 Confirm the LMC Operation To be stated...
Страница 120: ......
Страница 296: ...8 10...
Страница 300: ......
Страница 301: ......