19
18
5. Carefully proceed to close the back door with the calibration card inside
the machine.
6. Turn the machine face side to yourself and press START/STOP button to start
the calibration process.
7. Calibration will begin and the calibration screen will display a message
“CIS check wait”
8. When calibration is done “CIS check ok” message must be displayed on
the screen. If “CIS check fail” message is displayed, calibration card must be
readjusted and placed exactly on top of the CIS sensors and steps from 3 to 6
should be repeated.
9. When “CIS check ok” is displayed, remove the calibration card from the inside
of the machine.
10. Reset the machine by turning it off, then on.
11. Same steps must be taken for the UV + IR calibration process.
12. The CIS calibration is done.
6.4 Troubleshooting
NOTE:
The Domino, like all other discriminators on the market with intergrated
detectors, cannot guarantee a counted bill to be genuine or counterfeit. There are
several reasons a genuine bill may be recognized as counterfeit, or vice versa.
Possible reasons include: direct sunlight exposure, worn bills, excessively dirty
bills, or highly advanced counterfeiting techniques used to produce the bill.
Separate equipment may be required to definitively determine whether or not a bill
is counterfeit.