KV4000 Kinematic Viscosity Bath 

Operation and Technical Manual 

 

KV4000-Manual 

12 -

 

 

Cannon®-Fenske Opaque Reverse Flow 
Viscometers 

The reverse flow viscometers are designed for 

testing  opaque  liquids.  These  viscometers  wet 
the  timing  section  of  the  viscometer  capillary 
only  during  the  actual  measurement  and  must 
be  cleaned,  dried  and  refilled  before  a  repeat 
measurement  can  be  made.  By  contrast,  other 
viscometer  types  commonly  used  to  measure 
transparent  liquids  allow  the  sample  to  be 
repeatedly drawn up into the capillary, permitting 
duplicate measurements.  

 

  For measurement of transparent and dark 

liquids  having  kinematic  viscosities  of  up 
to 20,000cSt  

  Requires a sample of approx 12mL.  

  Can be used for viscosities of asphalts by 

ASTM D2170 method.  

 

Catalog No. 

Size 

Approximate 

Constant, 

cSt/s 

Kinematic 

Viscosity Range, 

cSt 

378-025-C02 

25 

0.002 

0.4 to 2 

378-050-C02 

50 

0.004 

0.8 to 4 

378-075-C02 

75 

0.008 

1.6 to 8 

378-100-C02 

100 

0.015 

3 to 15 

378-150-C02 

150 

0.035 

7 to 35 

378-200-C02 

200 

0.1 

20 to 100 

378-300-C02 

300 

0.25 

50 to 250 

378-350-C02 

350 

0.5 

100 to 500 

378-400-C02 

400 

1.2 

240 to 1,200 

378-450-C02 

450 

2.5 

500 to 2,500 

378-500-C02 

500 

1,600 to 8,000 

378-600-C02 

600 

20 

4,000 to 20,000 

378-650-C02 

650 

45 

9,000 to 45,000 

378-700-C02 

700 

100 

20,000 to 100,000 

 
Ubbelohde  

 

Ubbelohde  viscometers  measure  transparent 

liquids, and unlike the Cannon

®

-Fenske Routine 

viscometers, they maintain the same viscometer 
constant 

at 

all 

temperatures. 

This 

is 

advantageous  when  samples  are  to  be 
measured at different temperatures.  

 

 

Suspended-level  type  viscometers  are  for 
transparent liquids of up to 100,000cSt   

 

Requires  a  sample  volume  of  approx 
11mL.  

 

Catalog No. 

Size 

Approximate 

Constant,  

cSt/s 

Kinematic Viscosity 

Range, cSt 

378-000-C03 

0.001 

0.3 to 1 

378-00C-C03 

0C 

0.003 

0.6 to 3 

378-00B-C03 

0B 

0.005 

1 to 5 

378-001-C03 

0.01 

2 to 10 

378-01C-C03 

1C 

0.03 

6 to 30 

378-01B-C03 

1B 

0.05 

10 to 50 

378-002-C03 

0.1 

20 to 100 

378-02C-C03 

2C 

0.3 

60 to 300 

378-02B-C03 

2B 

0.5 

100 to 500 

378-003-C03 

200 to 1,000 

378-03C-C03 

3C 

600 to 3,000 

378-03B-C03 

3B 

1,000 to 5,000 

378-004-C03 

10 

2,000 to 10,000 

378-04C-C03 

4C 

30 

6,000 to 30,000 

378-04B-C03 

4B 

50 

10,000 to 50,000 

378-005-C03 

100 

20,000 to 100,000 

 
Viscometer Holders 

 

To  perform  viscosity  measurements,  Koehler 

offers  viscometer  holders  for  use  with  the 
KV4000.  The  correct  holder  must  be  used  with 
the  corresponding  viscometer  tube  for  proper 
operation: 

 

 

Viscometer Tube Type 

Corresponding Holder 

Cannon®-Fenske Routine 

K23381 

Reverse Flow 

K23383 

Ubbelohde 

K23382 

 

Catalog No. 

Size 

Approximate 

Constant, 

cSt/s 

Kinematic 

Viscosity Range, 

cSt 

378-025-C01 

25 

0.002 

0.5 to 2 

378-050-C01 

50 

0.004 

0.8 to 4 

378-075-C01 

75 

0.008 

1.6 to 8 

378-100-C01 

100 

0.015 

3 to 15 

378-150-C01 

150 

0.035 

7 to 35 

378-200-C01 

200 

0.1 

20 to 100 

378-300-C01 

300 

0.25 

50 to 250 

378-350-C01 

350 

0.5 

100 to 500 

378-400-C01 

400 

1.2 

240 to 1,200 

378-450-C01 

450 

2.5 

500 to 2,500 

378-500-C01 

500 

1,600 to 8,000 

378-600-C01 

600 

20 

4,000 to 20,000 

378-650-C01 

650 

45 

9,000 to 45,000 

378-700-C01 

700 

100 

20,000 to 100,000 

Содержание KV4000

Страница 1: ...re Avenue Bohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Anal...

Страница 2: ......

Страница 3: ...e with the applicable standards set forth in this certification Specifications ASTM D445 ASTM D2532 ASTM D6074 ASTM D6158 IP 71 ISO 3104 DIN 51550 FTM 791 305 NF T 60 100 This unit is tested before it...

Страница 4: ......

Страница 5: ...KV4000 Kinematic Viscosity Bath Operation and Instruction Manual KV4000 Manual 1...

Страница 6: ...KV4000 Kinematic Viscosity Bath Operation and Instruction Manual KV4000 Manual 2...

Страница 7: ...rate statement is given cataloging the items that require separation from the equipment along with basic information on subsequent processing or recycling prior to disposal of the equipment http www d...

Страница 8: ...1716 As a minimum the following substances preparations and components have to be removed from any separately collected WEEE Mercury containing components such as switches or backlighting lamps compac...

Страница 9: ...unning Tests 11 4 3 Temperature Controller Operation 13 5 Operation 14 5 1 Bath 14 5 2 Start Up 15 5 3 Control Panel Operation 15 5 4 Test Procedure 17 6 Safety Features 18 6 1 Over Temperature Protec...

Страница 10: ...cosity Bath Operation and Instruction Manual KV4000 Manual 6 9 Wiring Diagrams 22 9 1 K23702 K23708 115 V Wiring 22 9 2 K23792 K23798 220V Wiring 23 10 Service 24 11 Storage 24 12 Warranty 24 13 Retur...

Страница 11: ...rate and reliable instrumentation as well as support for your laboratory testing programs Please do not hesitate to contact us at any time with your inquiries about equipment tests or technical suppor...

Страница 12: ...for use in accordance with the applicable standard test methods listed in section 1 2 of this manual The use of this equipment in accordance with any other test procedures or for any other purpose is...

Страница 13: ...ment from cart and place on bench 8 Remove foam insert Borosilicate Glass Jar as per steps 9 thru 19 9 Using a flat head screwdriver remove the four 4 silver screws located at the four corners of the...

Страница 14: ...s to these during normal operation can shatter the glass components releasing the bath fluid posing thermal burn hazards and slipping hazards For this reason the instrument must be operated by and acc...

Страница 15: ...y is placed into the bath through the viscometer ports on top of the instrument 7 Bath Bath Stirrer The stirrer constantly circulates bath medium to prevent temperature gradients and ensures temperatu...

Страница 16: ...nt at all temperatures This is advantageous when samples are to be measured at different temperatures Suspended level type viscometers are for transparent liquids of up to 100 000cSt Requires a sample...

Страница 17: ...ase the set point temperature and to increase or change parameter values when programming the temperature controller 6 Exit Function Key This key is used to exit or leave a level One level backward IM...

Страница 18: ...ed for performing kinematic viscosity tests using glass capillary viscometers Please be sure to read the safety and hazard warnings the installation procedure and any of the standard test methods befo...

Страница 19: ...matches that of the ASTM Thermometer Calibrate if necessary See Section 4 3 for full operational details of the temperature controller 5 3 Control Panel Operation Figure 3 Control Panel_Main Screen 1...

Страница 20: ...an be found on the calibration certificate provided with the viscometer 5 Constant 2 Press the Constant 2 button to input a second viscometer tube constant value using the pop up numerical keypad A se...

Страница 21: ...uum system to the initial testing position above the first line This is also called charging the viscometer Place stopper in the top of the viscometer tube and remove the withdrawal bulb NOTE The proc...

Страница 22: ...KV4000 Kinematic Viscosity System is equipped with several safety and protection features which are described in the following sections WARNING The bath is NOT Explosion Proof 6 1 Over Temperature Pro...

Страница 23: ...tenance WARNING Disconnect power to the unit before servicing to avoid exposure to high voltages and or temperatures which may result in personal injury or death If you have any questions about mainta...

Страница 24: ...move the four 4 silver screws located at the four corners of the back panel 7 Remove the back panel plate and set aside 8 A collection of wires runs from the lower control panel up the left side while...

Страница 25: ...lacement Parts Part Number Description 050 001 028 Single Pole Switch 15A 091 032 001 Solid State Relay 4 32 VDC 20A 275 103 036 RTD Temperature Safety Limit 120V 275 103 037 RTD Temperature Safety Li...

Страница 26: ...KV4000 Kinematic Viscosity Bath Operation and Technical Manual KV4000 Manual 22 9 Wiring Diagrams 9 1 K23702 K23708 115 V Wiring...

Страница 27: ...KV4000 Kinematic Viscosity Bath Operation and Technical Manual KV4000 Manual 23 9 2 K23792 K23798 220V Wiring...

Страница 28: ...rn of the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive remedy for any claim arising out of the purchase of any product is the repair or re...

Страница 29: ...KV4000 Kinematic Viscosity Bath Operation and Technical Manual KV4000 Manual 25 Notes...

Страница 30: ...KV4000 Kinematic Viscosity Bath Operation and Technical Manual KV4000 Manual 26 Notes...

Отзывы: