During a test, the regulator follows
five stages:
1. Hold time of ten minutes at the
oven temperature at the start of
the test. Entry of nitrogen to the
oven (high flow rate).
2. Heating according to a ramp of
12.5° C/min up to 500° C.
Entry of nitrogen in the oven (low
flow rate).
3. Hold time of fifteen minutes at
500° C.
Entry of nitrogen in the oven (low flow rate).
4. Cooling with compressed air for forty minutes. Entry of nitrogen in the oven (high
flow rate).
5. End of test. Switch off heating and gas supply.
During the test, check that the nitrogen flow rates are correct. For
guidance, for a flow rate of 150 ml/min, the flow meter should be set at
20.62 mm and for a flow rate of 600 ml/min, at 58.20 mm. Always refer to the
abacus received with the appliance.
When the test is completed, the RUN LED will flash. ‘
END
’ is indicated on
the lower display. The temperature regulator will remain in this status until it
is reinitialized.