K30160, K30165, K30166 Rust Preventing Characteristics Oil Bath 

Operation and Instruction Manual 

K30160,5,6-Manual

 

-12-

 

4  Descriptions 

 

4.1  Temperature Controller Operation 

 

 

 
 

Figure 4. Temperature Controller 

 

1.  Process  Temperature  Display.

  The 

upper red LED display shows the process 
temperature as read from the RTD probe. 

 

2.  Set  Point  Temperature  Display.

  The 

lower  green  LED  display  shows  the  set 
point temperature of the controller. 
 

3.  Programming  Key.

  Permits  scrolling 

through  controller  menu  parameters.  One 
Level Forward 
 

4.  Down  Key.

  Used  to  decrease  the  set 

point  temperature  and  to  decrease  or 
change 

parameter 

values 

when 

programming the temperature controller.  
 

5.  Up  Key.

  Used  to  increase  the  set  point 

temperature  and  to  increase  or  change 
parameter  values  when  programming  the 
temperature controller. 
 

6.  Exit  /  Function  Key.

  This  key  is  used  to 

exit or leave a level. One level backward  
 

IMPORTANT 

NOTE:

 

The 

digital 

temperature  controller  for  the  unit  comes 
pre-programmed from the Koehler factory. 
Please do NOT attempt to re-program the 
digital  temperature  controller  as  this  will 
void the product  warranty.  If assistance is 

required, please do not hesitate to contact 
the Koehler technical service department. 

 

Setting  the  Temperature

.  Set  the  desired 

operating  temperature  by  adjusting  the  set 
point with the up and down keys. The set point 
will  be  displayed  in  the  lower  green  Set  Point 
LED display and the actual temperature will be 
displayed  in  the  upper  red  Process  LED 
display.  Please  allow  the  instrument  to  fully 
equilibrate before proceeding with any testing. 

 

Temperature Calibration.

  This routine allows the 

digital  temperature  controller  to  be  calibrated 
to a certified thermometer. 

 

a.  Use  a  certified  calibrated  measuring 

device  to 

acquire  the 

temperature. 

Calculate  the  difference  between  the 
measuring  device  and  the  Process  value 
displayed on the controller. 

b.  Press the program key two times until 

PCt

 

is  displayed  in  the  lower  green  LED 
display.  Press  the  DOWN  key.  CAL  will 
display on the lower green display. If there 
is  a  value  observed  in  the  upper  red  LED 
display, add  it  to the calculated difference 
obtained  in  the  previous  step.  This  is  the 
offset value. 

 

c.  Press  the  Program  Key.  The  lower  green 

display will flash. Use the up or down keys 
to  adjust  to  the  new  calibration  offset 
value  on  the  upper  red  display  calculated 
in  the  previous  step.  When  the  value  has 
been 

entered, 

the 

controller 

will 

automatically  store  the  value.  The  lower 
green  display  will  stop  flashing.  If  further 
adjustments  are  necessary,  press  the 
Program  Key  again.  Resume  regular 
operations by pressing  the Exit / Function 
key two times. Verify if the new calibration 
is  correct  by  observing  the  upper  red 
display  and  comparing  the  value  with  the 
calibrated reference device.  

 
 

Auto  Tune.

  This  routine  allows  the  digital 

temperature  control  to  learn  the  heating 
parameters needed for any particular set point 
temperature.  This  operation  should  be  done 
when  installing  a  new  unit,  after  replacing  or 
changing  the  bath  medium  type,  or  utilizing  a 
different  temperature  set  point  20%  different 
from 

the 

previously 

used 

set 

point 

temperature. 

Содержание K30160

Страница 1: ...ia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis Instrument...

Страница 2: ......

Страница 3: ...istics was manufactured in conformance with the applicable standards set forth in this certification Specifications ASTM D665 ASTM D3603 ASTM D6158 IP 135 ISO 7120 DIN 51585 FTM 791 4011 NF T 60 151 T...

Страница 4: ...K30160 K30165 K30166 Rust Preventing Characteristics Oil Bath Operation and Instruction Manual K30160 5 6 Manual 4...

Страница 5: ...K30160 K30165 K30166 Rust Preventing Characteristics Oil Bath Operation and Instruction Manual K30160 5 6 Manual 5...

Страница 6: ...with basic information on subsequent processing or recycling prior to disposal of the equipment http www dpa system dk en WEEE Products Private or professional use Responsibility for Registration and...

Страница 7: ...any separately collected WEEE Mercury containing components such as switches or backlighting lamps compact fluorescent lamps CFL Batteries Printed circuit boards if the surface of the printed circuit...

Страница 8: ...and Warnings 10 3 Getting Started 11 3 1 Packing List 11 3 2 Unpacking 11 4 Descriptions 12 4 1 Temperature Controller Operation 12 5 Operation 13 6 Safety Features 13 6 1 Over Temperature Protection...

Страница 9: ...or a full range of accurate and reliable instrumentation as well as support for your laboratory testing programs Please do not hesitate to contact us at any time with your inquiries about equipment te...

Страница 10: ...mended and may be extremely hazardous Chemical Reagents Information Chemicals and reagents used in performing the test may exhibit potential hazards Any user must be familiarized with the possible dan...

Страница 11: ...ation 5 Ensure that all parts listed on the packing list are present Inspect the unit and all accessories for damage If any damage is found keep all packing materials and immediately report the damage...

Страница 12: ...Process LED display Please allow the instrument to fully equilibrate before proceeding with any testing Temperature Calibration This routine allows the digital temperature controller to be calibrated...

Страница 13: ...WARNING There are exposed hot surfaces on the bath Avoid contact with exposed skin by the use of protective equipment as required or serious injury could occur 5 Test Sample Setup Instructions Fill th...

Страница 14: ...s all painted surfaces and glass either a solution of soap and water or laboratory grade detergent may be used Apply cleaner to clean wipe or cloth not to the instrument directly Wipe surface clean Do...

Страница 15: ...hould not be subjected to extremes of temperature and or moisture This equipment was shipped from the factory in a corrugated cardboard container If long term storage is anticipated re packing the ins...

Страница 16: ...ntainer when you return the material to our plant Shipping containers without an RA number prominently displayed with will be returned to the sender Goods must be returned freight prepaid Returns will...

Страница 17: ...K30160 K30165 K30166 Rust Preventing Characteristics Oil Bath Operation and Instruction Manual K30160 5 6 Manual 17 13 Assembly and Wiring Diagrams...

Страница 18: ...K30160 K30165 K30166 Rust Preventing Characteristics Oil Bath Operation and Instruction Manual K30160 5 6 Manual 18...

Страница 19: ...K30160 K30165 K30166 Rust Preventing Characteristics Oil Bath Operation and Instruction Manual K30160 5 6 Manual 19...

Страница 20: ...K30160 K30165 K30166 Rust Preventing Characteristics Oil Bath Operation and Instruction Manual K30160 5 6 Manual 20...

Страница 21: ...K30160 K30165 K30166 Rust Preventing Characteristics Oil Bath Operation and Instruction Manual K30160 5 6 Manual 21 Notes...

Страница 22: ...K30160 K30165 K30166 Rust Preventing Characteristics Oil Bath Operation and Instruction Manual K30160 5 6 Manual 22 Notes...

Страница 23: ...K30160 K30165 K30166 Rust Preventing Characteristics Oil Bath Operation and Instruction Manual K30160 5 6 Manual 23 Notes...

Отзывы: