K30160, K30165, K30166 Rust Preventing Characteristics Oil Bath
Operation and Instruction Manual
K30160,5,6-Manual
-11-
3 Getting Started
3.1 Packing List
Component Check List:
K30160, K30165, or K30166 Rust Preventing
Characteristics Oil Bath
K30100 Steel Test Specimens with Type 2
Plastic Specimen Holders (6)
K30120 Plastic Beaker Covers (6)
Accessories (purchased separately)
332-002-006 Test Beaker, 400mL (6)
250-000-09F ASTM 9F Thermometer
Range: 20 to 230°F
250-000-09C ASTM 9C Thermometer
Range: -5 to +110°C
K30130 Chuck for Polishing Test Specimens
380-150-002 Aluminum Oxide Cloth, 150-grit,
pack of 50
380-240-002 Aluminum Oxide Cloth, 240-grit,
pack of 50
K30140 Auxiliary Stirrer Blade
K30110 Steel Test Specimen without holder
3.2 Unpacking
1.
This unit comes packed in a crate with the rust
test specimens packed in a separate carton
inside the main crate.
2.
Extract instrument and place on suitable cart
for transportation to work area / lab bench.
WARNING:
Be sure two or more individuals
are available for extracting and lifting instrument
from box to cart and from cart to bench.
Individuals must lift in accordance to proper
technique. See Figure below.
3.
Lift instrument from cart and place on bench.
4.
Unpack the rust test specimens from the
separate carton and place in a safe location.
5.
Ensure that all parts listed on the packing list
are present. Inspect the unit and all
accessories for damage. If any damage is
found, keep all packing materials and
immediately report the damage to the carrier.
We will assist you with your claim, if
requested. When submitting a claim for
shipping damage, request that the carrier
inspect the shipping container and equipment.
Do not return goods to Koehler without written
authorization.
Environmental Conditions:
The instrument
environment must comply with the following
conditions for proper setup:
No / Low Dust
No direct sunlight
Not near heating or AC ventilation ducts
No Vibrations
Clearance from other instruments
Temperature Range: 5 to 40°C
Elevation to 2000 meters
Relative Humidity: < 80%
Ventilation.
A fume hood or exhaust system is
required when operating the unit. Flammable
vapors and/or steam are generated during
operation and must not be permitted to
accumulate. A canopy-style hood may be used if
the height from the top of the unit to the canopy is
5 feet or less. The exhaust blower should have a
rating of 1000 C.F.M. or greater.
Power.
Connect the line cords to properly fused
and grounded receptacles with the correct voltage
as indicated in section 1.3 or on the back of the
unit.
WARNING:
For safety, disconnect the
power when performing any maintenance and/or
cleaning.
Do
NOT
turn the power on unless the
bath is filled with the proper medium; otherwise,
damage may occur to the unit and the warranty
will be void.