Diagnostics
9B8949 – 10SEP02
75
Checking the FIXER PUMPS
[1]
Enter mode 5. See
Caution
Operating the PUMPS excessively without solution
might cause damage to the PUMPS.
[2]
Press and hold the “RUN” BUTTON on the
CONTROL BOARD.
Note
Holding the “RUN” BUTTON:
• Actuates the FIXER RECIRCULATION PUMP and
illuminates LED 3
• Actuates the FIXER REPLENISHMENT PUMP and
illuminates LED 5
[3]
Do LED 3 and LED 5 illuminate?
[4]
If LED 3 and LED 5 illuminate and the PUMPS do not operate, check FUSE F1 on the DRIVE BOARD.
[5]
Does FUSE F1 have continuity?
[6]
Check the output signal for the FIXER PUMP CIRCUITS on the DRIVE BOARD:
[7]
Are the voltages correct?
H164_0034AC
FIX.
DEV.
RUN
Rep.
Dev. Temperature
Low
Ready
High
Service
Yes
No
Continue with
a. Check:
• I/O BOARD
• I/O BOARD/DRIVE BOARD CABLE
b. If necessary, install a new part.
Yes
No
Continue with
a. Install a new FUSE.
b. Continue with
Signal
Position on DRIVE BOARD
Voltage
FIXER RECIRCULATION PUMP
PIN 5 to PIN 17 on CN2
approximately 100 V AC
FIXER REPLENISHMENT PUMP
PIN 21 to PIN 9 on CN2
approximately 100 V AC
Yes
No
a. Check the PUMP.
b. If necessary, install a new PUMP.
a. Check:
• SOLID STATE RELAY SSR3
• SOLID STATE RELAY SSR5
b. If necessary, install a new SOLID STATE
RELAY.