Acquiring a Panoramic, TMJ x2 or Sinus Image
KODAK 9000 3D Extraoral Imaging System_User Guide (SM710)_Ed 01
5–3
5. On the
Control Panel
click
to turn ON the positioning laser beams. Press and hold
to correctly position the patient using the 2 positioning laser beams:
•
The mid-sagittal (1) positioning laser beam for a vertical alignment.
•
The horizontal (2) positioning laser beam for a Frankfurt plane alignment.
6. Immobilize the patient head with the temple supports
(3).
NOTE
You can re-activate the laser positioning beams as needed. You can
press the same button to turn OFF the laser beams, or wait 60 seconds
for the beams to turn OFF automatically.
IMPORTANT
The spinal column and the nose of the patient must be aligned in a
straight line (1).
2
1
2
1
2
1
3