DIAGNOSTICS
Status Messages and Error Codes
30JUL07
8F2919
Page
217 of 326
Error 27-932: OPTICS Communications Retry
Messages
Description
This error occurs when the MCS tries to communicate with the MICROCONTROLLER but
receives no response.
Note
Error 27-931 is generated when this problem is detected after 5 successive communications
attempts.
Result of Error
None
User Action
None
FE Action
None, this is a log notification only.
LOCAL PANEL
LOG
DICOM
SEVERITY
None
(No alarm)
MCS failed to communicate
with Optics (BPM)
NORMAL
3-Notify