Using the KODAK EKTACOLOR Processing Cartridge 111 With Digital Minilabs, CP-48S/CP-49E

 

 

 CIS-229E

 

10 

 

The Black Patch monitors the performance of the 
developer for activity and contamination. There is 
no upper control limit for Black Patch. If the Black 
Patch plots are low, then it could be an indication of 
low developer activity due to underreplenishment 
or severe oxidation. If the blue Black Patch is very 
low and separated from the red and green plots, it 
could be a sign of developer contamination. See 
prescriptions listed under the LD parameter. 

 

The HD-LD (Contrast) is usually a very steady plot 
parameter that is unresponsive to most process 
problems. However, it can help monitor developer 
activity problems due to agitation or oxidation.  If 
developer tank solution is underagitated, the plots 
will be low. If developer tank solution is slightly 
oxidized, the plots will be high. If the developer 
becomes extremely oxidized, the plots will be low. If 
the solution is extremely out of control by more 
than 0.18density units over or under aim, drain and 
replace the developer tank solution. If the plots are 
less than 0.18 density units over or under aim, a 
prescription can be risked 
Prescriptions: 

— Underagitation: call service representative to 

fix agitation pump on processor. 

— Oxidation: check for air bubbles in developer 

tank; call service technician if seen. 

 

 

The LD (Speed) monitors the developer activity. 
Speed is the primary indicator of developer time, 
temperature, and especially developer 
replenishment rate. For time that is too long, 
temperature too high, or overreplenishment, the LD 
plots will be high. For time too short, temperature 
too low, or underreplenishment, the LD plots will be 
low. Check the temperature of the developer; adjust 
as necessary to bring back into specification. If the 
plots are extremely out of control, by more than 
0.18 density units over or under aim, drain and 
replace the developer. If the plots are less than 0.18 
density units from aim, a prescription can be risked.  
 

Note: If the red LD is very high and/or the green LD is 
very low and they are split from the blue LD, it may be 
caused by developer contaminated with bleach-fix. In 
this case, the developer must be drained and replaced 
with fresh developer tank solution. Make sure the tank 
is cleaned and thoroughly flushed with water to 
remove all contaminated developer from the tank and 
circulation system. 

 

 
 

Prescriptions: 

— For high LD plots:  

Overreplenishment: Remove a volume of 
developer solution equal to 10% of the total 
tank volume and replace with water. Repeat 
until it is in control. 

— For low LD plots:  

Underreplenishment: Remove a volume of 
developer tank solution equal to one third of 
the tank volume and replace with fresh 
developer tank that is mixed according to the  
“Make 1 litre” component ratios in Table 1a. 

 

 

D-min (Stain) monitors the whiteness of the 
unexposed paper. An increase in the D-min plot can 
indicate oxidation or contamination problems in the 
developer, or a problem in the super rinse when the 
solution is dirty or has biological growth residue. 

 

Prescriptions: 

— For high D-min on paper caused by oxidized or 

contaminated developer, replace the developer 
tank solution with fresh developer tank solution 
mixed according to the component ratios in 
Table 1a. 

— For a high D-min on the paper caused by dirty 

stabilizer or biological growth in the stabilizer 
tanks and racks, clean the stabilizer tanks and 
replace the solution with water and KODAK 
Rinse Tablets.

 

Содержание CP-48S

Страница 1: ...will produce prints with rich bright colors and excellent shadow detail The KODAK EKTACOLOR Processing Cartridge 111 is compatible with both FUJI and NORITSU Minilab Systems using Processes CP 48S an...

Страница 2: ...ridge However if you are converting from a FUJI CP 48LR or CP 49ELR Processing Cartridge you will need to adjust the replenishment rates for the Developer Replenisher and Bleach Fix Replenisher to the...

Страница 3: ...er Developer P1 Bleach Fix P2 Super Rinse PS4 Super Rinse PS3 Super Rinse PS2 Super Rinse PS1 22 27 22 27 21 45 total 35 95 101 107 42 0 3 113 0 0 5 40 0 2 0 104 0 3 0 The P2 bleach fix and PS 1 super...

Страница 4: ...ter Developer P1 Bleach Fix P2 Super Rinse PS4 Super Rinse PS3 Super Rinse PS2 Super Rinse PS1 19 22 19 27 17 32 43 46 109 115 43 0 0 3 109 4 0 5 43 0 0 2 109 4 3 0 The P2 bleach fix and PS 1 super ri...

Страница 5: ...IME SP Developer Replenisher LORR To the specified volume of water add the PRIME SP Developer Replenisher LORR concentrate directly from the bottle then EKTACOLOR RA Developer Starter If necessary add...

Страница 6: ...blet to each stabilizer tank and fill the tank with water to the appropriate level DISPOSING OF EMPTY CARTRIDGES Your processor rinses the chemical solutions from the cartridge during the mixing cycle...

Страница 7: ...for FUJI FRONTIERS only It is necessary to add water to the developer replenisher tank Add this instruction to those above 2A Pour 500 mL of water into the developer the slot farthest to the left in...

Страница 8: ...EDGE Papers For calculating the aims for your control strip code read the Status A Red Green and Blue densities of the reference strip and then apply the correction factors For the HD LD parameter su...

Страница 9: ...ontrol Consider that you may have more than one problem occurring at the same time Consider each potential cause of the out of control condition and verify the operational conditions of the processor...

Страница 10: ...mperature too low or underreplenishment the LD plots will be low Check the temperature of the developer adjust as necessary to bring back into specification If the plots are extremely out of control b...

Страница 11: ...s ol s ol s ntro t o ntro ntro con co con con c c c c c a r r p p s o n c d d d d bed bed bed b d crib crib crib ib esc esc esc de de de d se se se au au au e ca e ca e ca the the the th or t or t or...

Страница 12: ...HESTER NY 14650 Using the KODAK EKTACOLOR Processing Cartridge 111 with Digital Minilabs Using Processes CP 48S and CP 49E KODAK Publication No CIS 229E Kodak Edge Ektacolor Prime and Royal are tradem...

Отзывы: