7. Mechanical 

Connection 

Before installation: 

  Make sure that the permitted maximum operating pressure and the operating 

temperature of the device are not exceeded. 

  Make sure that the electrical supply for the devices corresponds with the 

operational data of the device. 

  Make sure that no remains of packing material exist within the device. 

  The devices should not be installed within an inductive field. 

  If possible, it should be checked, after the mechanical installation whether the 

connection joints/ piping is properly sealed. 

 

Attention! A displacement of switching range occurs if the flow takes 
place from top to bottom. 

 
 

 

8. Electrical 

Connection 

Attention! Make sure that the voltage values of your system 
correspond with the voltage values of the measuring unit, mentioned 
on type-label. 

 

 

  Make sure that the supply wires are de-energized. 

  The connection of the bistable reed contact succeeds is accomplished through 

the cable on the contact housing. The potential-free contact closes on rising 
level above the adjusted flow value.  

 
 
 

Содержание SWK 1101

Страница 1: ...Operating Instructions for Float Flow Meter Monitor Model SWK...

Страница 2: ...onnection 6 8 Electrical Connection 6 9 Operation Configuration Adjustments 7 9 1 Adjustment of Limit Value SWK 1 7 9 2 Adjustment of Limit Value SWK 2 7 10 Maintenance 8 11 Technical Information 8 12...

Страница 3: ...hen the machines fulfil the EC machine guidelines as per PED 2014 68 EU In acc with Article 4 Paragraph 3 Sound Engineering Practice of the PED 2014 68 EU no CE mark Diagram 8 Pipe Group 1 dangerous f...

Страница 4: ...C cable Brass version Attached reed contact N O with 2m PVC cable PVC VA version Attached and aligned pointer display option 5 Operating Principle KOBOLD Flow Meter Monitor of type SWK operates accord...

Страница 5: ...IIA IIB and IIC b In the Zone 22 Dust Ex Category 3D at non conductive dusts with a minimum igniting energy of 3 mJ c In the Zone 1 Gas Ex Cat 2G into explosion group of IIA IIB and IIC d In the Zone...

Страница 6: ...If possible it should be checked after the mechanical installation whether the connection joints piping is properly sealed Attention A displacement of switching range occurs if the flow takes place f...

Страница 7: ...line and let the flow medium enter slowly till the pointer display shows the desired flow value at which the rising flow should operate the switch the value for the decreasing flow lies below due to t...

Страница 8: ...ainless steel 1 4310 Gaskets none Butt rings SWK x1 brass Ms 58 SWK x2 stainless steel 1 4301 SWK x13 PVC Max temperature SWK x1 SWK x2 100 C SWK 13 60 C Max pressure SWK x1 SWK x2 250 bar SWK 13 6 ba...

Страница 9: ...ax load 5 107 switching operations at low load 10V 1mA 108 switching operations 12 Order Codes Example SWK 1101 Switching range L min water Connection female Flow monitors Flow meters and monitors Swi...

Страница 10: ...SWK page 10 SWK K06 0618 13 Dimensions...

Страница 11: ...aration relates is in conformity with the standards noted below EN 61010 1 2011 Safety requirements for electrical equipment for measurement control and laboratory use EN 60529 2014 Protection through...

Отзывы: