8.7 Additional parameters of the flow metering
P76: Head of flow (LEV) (Read only parameter)
The Headwater value can be checked here. This is the “h” value in the formula for
flow calculation.
P77: TOT1 volume flow totalised (resettable)
P78: TOT2 volume flow totalised (non-resettable)
Supplementary parameter of the logger
P79: Free space of logger in percent
If the value is 0, the registry has overflown and every new entry will overwrite the
oldest one.
Clearing the logger
1). Enter parameter
P79
.
2). Press + keys.
3). The display flashes „Lo-Clr” message.
4). Pressing
E
will clear the logger.
Other parameters
P96: Software code 1 (Read only parameter)
P97: Software code 2 (Read only parameter)
P98: Hardware code (Read only parameter)
P99: dcba Access lock by secret code
The purpose of this feature is to provide protection against accidental programming
or intentional reprogramming of parameters by a person not entitled to do so. The
secret code can be any value other than
0000
. Setting a secret code will
automatically be activated when the NUS-7 is returned to the Measurement Mode.
In order to program locked device the secret code should be entered first in
P99.
Thus for entering a new code or erasing the old one the knowledge of the previous
code is necessary.