Transmitted value
VOL
f
P40…P45
(H-DIST)
VOL%
=
P10
m
X
H
P10
P11
*
VOL
Parameters to set
P00
P01(a) =
3
P02(b)
P04 = H
P05
≥
X
m
P40…P45
P00
P01(a) =
4
P02(b)
P04 = H
P05
≥
X
m
P10 = X
0%
P11 = X
100%
P40…P45
A:
Shortest measurable distance
B:
Volume (content) pertaining to the greatest measurable level
C:
Whole value of the vessel
D:
diagram valid for the default value of P10 P11
P02:- c b a Calculation units
FACTORY DEFAULT: 000
a
Temperature
0
°C
1
°F
This table is interpreted according to
P00(c)
,
P01(a)
and
P02(c)
and is irrelevant in
case of percentage measurement [
P01(a)= 2
or
4
)]
c
Time
Attention!
NUS-7 is a level transmitter.
Although it can be used for
measuring weight, due to factors
involved in doing so, accuracy
may essentially be influenced.
0
s
1
min
2
hour
3
day
P04:-Maximum Distance to be Measured (H)
FACTORY DEFAULT: X
M
as per chart
This is the only parameter that has to be programmed for each application
other than distance (however to avoid disturbing effect of possible multiple
echoes it is suggested to do this in distance measurement applications too).
b
Volume
Weight (set also P32)
Volume flow
Metric
US
Metric
US
Metric
US
0
m
3
ft
3
-
lb
(pound)
m
3
/time ft
3
/time
1
litre gallon
tons tonnes
litre/time
gallon/
time