Kobold NRF Скачать руководство пользователя страница 5

aran_v02 or higher

NRF-100_KM Rev. : 1.0

16 / 04 / 2002

01 / 12 / 2002

Output :

4 - 20 mA , 2-wire

Maximum Loop
Resistance :

350

at 12VDC, 950

at 24VDC

Calibration :

Via 4 push-button switches

Capacitance range :

10 pF to 10000 pF

Resolution :

±0.1%

Linearity Due to
Temperature :

Negligible (less than 100 ppm)

Minimum Span :

10 pF

Accuracy and
Repeatability :

±1% of span or less

Output Damping :

One turn pot, 0 - 10 sec. delay

Sensing System :

Operating Temperature: Electronics -40° to 70°C

Capacitance

Maximum Probe
Frequency :

1.2 Mhz

Maximum Power at :

8 mW

Hazardous Areas:

Consult Factory

Power :

12 VDC - 36 VDC

Electrical Specifications :

Mechanical Specifications :

Enclosures

Standard :

NEMA 4 (

Polyamide

)

316 SS Fitting :

500 PSI @ 70°F, 250 PSI @ 300°F
100 PSI @ 350°F

Tri-Clamp :

Per tri-clamp rating

KOBOLD INSTRUMENTS, INC.

1801, Parkway View Drive, Pittsburgh, PA 15205

Telephone: (412) 788-2830 • FAX : (412) 788-4890

Web site : www.koboldusa.com

Example:  L = 12” (30 cm) from the bottom of the tank

H1 = 72” (183 cm) from the bottom of the tank
H  = 96” (244 cm) from the bottom of the tank

( 72  -  12 )

mA  =  --------------   x  16  +  4  =  15.43

( 96  -  12 )

The correct loop current is is 15.43mA.

Depress SPAN S-UP or S-DW buttons as required to
obtain a meter reading of 15.43mA. If 15.43mA reading
has been obtained, the calibration is complete. 

Note: If a 15.43mA value cannot be reached, then
peform OFFSET AND RESET functions and re-start the
calibration

KOBOLD INSTRUMENTS CANADA, INC.

9A de l’Aviation, Pointe-Claire, Qc H9R 4Z2

Telephone : (514) 428-8090 • FAX: (514) 428-8899

Web site : www.kobold.com

Page 5

Optional:

St. Steel
Aluminum w/Epoxy Paint

Process Connection :

Many styles available, 3/4" NPT
standard

Maximum Pressure:

PTFE Fitting : 

150 PSI @ 70°F, 0 PSI @ 300°F

Отзывы: