Kobold NRF Скачать руководство пользователя страница 2

aran_v02 or higher

NRF-100_KM Rev. : 1.0

16 / 04 / 2002

01 / 12 / 2002

In order to calibrate the transmitter, you must use the loop
current meter. It should read currents in the range of 1.00
to 25.00 mA, with a resolution of .01 mA. Using a meter
of less resolution will somewhat reduce the calibration
accuracy. To calibrate the instrument :

1.

Remove enclosure (head) cover.

2.

Connect the loop current meter as per instructions in
WIRING section(Fig. 3).

3.

The loop current should now be in the range of 1.5
mA to 38 mA, which is normal at this point.

Three calibration procedures are described. Follow the
one which fits your application. Note the following defini-
tions used in the calibration procedures, referring to Fig 4. 

= the level of material which corresponds to 4.00 mA of 

loop current, i.e., the 0 %

L1 = a material level higher than L

H1 = a material level higher than L1, but less than H.

= the level of material in the vessel which corres-

ponds to 20.00 mA of loop current, i.e., the 100% level.

Note: To avoid the possibility of a “dead zone”, L must be
at least two ( 2 ) inches above the end of the probe for ( 4)
conductive  media  and  four  inches  above  for  non-
conductive media.

CALIBRATION L- H = when material in tank can be set
to L ( 0% ) and H (100 % ).

CALIBRATION L - H1 = when material in tank can be set
to L (0%) and H1 (less than 100 % ).

CALIBRATION L1 - H1 = when material in tank can be
set to L1 (greater than 0 % ) and H1 (less than 100 % ).

NOTE: Calibration procedure L-H gives the most 
accurate results and is the recommended procedure in 
all cases.

All wiring between the power supply and the transmitter
should be done with 18 AWG to 22 AWG shielded twisted
pair. The connection is made at the terminal strip within
the transmitter enclosure.

CAUTION: Units are designed to operate on the 12 to
36 VDC power only. Application of 110 VAC will destroy
the instrument.

1. Make sure the power source is turned off.

2. Pull power supply wires through conduit connection.

3. Connect the positive supply wire to the ( + ) terminal,

and the negative supply to the ( - ) terminal.

Note: Leave shield unattached at transmitter.

Connect the shield to ground at the power source.

4.

Replace the transmitter enclosure (head) cover until 

time to calibrate.

5.

Connect positive supply wire to the positive terminal

of the transmitter. See Fig. 3.

6.

Connect the loop current meter in series with the

negative supply wire as follows:

a. Negative transmitter wire to positive meter

terminal. See Fig. 3.

b. Negative meter terminal to negative power 

source termianl. See Fig.3.

7.

Turn ON the power. The meter may read anywhere
on the scale at either end. This is normal until calibra
tion has been completed. Proceed to the calibration 
Instructions.

Fig. 3

Page 2

Отзывы: