9. Adjustment
Possible adjustments of the device:
1. zero point
Prior to commissioning or at regular intervals, the zero point must be set in the
pressureless state, even if the display shows "0". Drifting into the minus range
cannot be ruled out despite the zero display.
Instructions for zero point setting on inside battery cover
2. password (factory pre-set: 5)
3. peak value memory
4. relay (option) with settable hysteresis
5. (factory pre-set: switching point on 50% of measuring range)
Factory pre-sets:
for MAN-LD
1. sampling rate (default: 5 measurements per second)
2. analogue output linear zum Messbereich 4...20 mA (option)
for MAN-SD
1. battery symbol on: voltage under 6,5 V
2. switch-off delay (default: 0 = inactive)
3. sampling rate (default: 5 measurements per second)
4. analogue output (linear) within measuring range 0…2 VDC (option)