![Kobold HND-C110 Скачать руководство пользователя страница 15](http://html1.mh-extra.com/html/kobold/hnd-c110/hnd-c110_operating-instructions-manual_1982440015.webp)
HND-C110
HND-C110 K05/0422
page 15
- Channel 1: current measuring value Cond, rES, TDS oder SAL (base address)
- Channel 2: temperature value
The measuring- and display range values read back from the interface are
always in the selected measurement unit!
12. Adjustment of temperature input
The temperature input can be adjusted with offset and scale. A reasonable
adjustment presumes reliable references (e.g. ice water, controlled precision
water bath, etc.).
If the inputs are adjusted (i.e. offset and scale are different from default settings)
the device will shortly display “Corr” after turned on.
Default setting for offset and scale are ‘off’ = 0.0, i.e. inputs are not changed.
Zero point correction:
Displayed value = measured value – OFFS
Zero point and slope correction: Displayed value = (measured value – OFFS) *
(1 + SCAL / 100)
Displayed value °F = (meas. value °F - 32°F - OFFS) • (1 + SCAL / 100)