HND-C110
page 14
HND-C110 K05/0422
Menu
Parameter
Value
Description
or
Set Output: Configuration of universal output
oFF
SEr:
Output off -> minimal power consumption
Serial interface activated
0,1,11…91
Base adress for serial interface communication
Set Corr: Measurement correction
Zero point adjustment/offset of temperature measurement
oFF
-5.0…5.0 °C
No zero point adjustment for temperature
measurement
Offset of temperature measurement in °C
Slope adjustment of temperature measurement
oFF
-5.00…5.00
No slope adjustment for temperature measurement
Slope correction of temperature measurement in [%]
Set Clock: Set real time
HH:MM
Clock
: set time hours:minutes
YYYY
Year
: set year
TT.MM
Date
: Set date day.month
rEAD CAL: Read calibration data:
see chapter 14.2 Calibration storage (rEAd CAL)
11. Universal output
If none of both is needed, we suggest to switch the output off, because battery life
then is extended.
11.1 The Serial Interface
All measuring and setting data of the device can be read and changed by means
of the serial interface and a suitable electrically isolated interface adapter
(HND-Z031). In order to avoid transmission errors, there are several security
checks implemented.
The following
standard software packages
are available for data transfer:
-
BUS-S20M:
20-channel software to simultaneously display
conductivity (channel 1)
temperature (channel 2)
resistivity (channel 3)
TDS (channel 4) and/or
salinity (channel 5).
If you want to use the interface functions make sure to switch off the
Auto-Range function.
The device has 2 channels: