TM/TMU/TM-SH Supplement Manual 
For Explosion Flow-meters 

Page 11 of 20 

 

3.5.4

 

Ambient temperature range Ta 

The ambient temperature range is dependant on the type of connection, installation, process 
temperature and temperature class. 
For determining the prevailing temperature class at the location of installation refer to the following 
tables as well as the relevant type certificate:  
With HAN R23 plug 

Neck extension (Between the 
TM, TMU or TM-SH

*

 sensors, 

and the Harting HAN R23 
Connector) 

Process temp. (°C) 

-50 °C to .. 

Ambient temp. 

range (°C) 

-40 °C to .. 

Temperature 

class 

without 

40 

45 

T6 

without 

60 

60 

T5 

  60 mm 

100 

80 

T4 

160 mm 

115 

80 

T4 

160 mm 

175 

80 

T3 

160 mm 

180 

80 

T2 

260 mm 

220 

80 

T2 

CAUTION: The temperature ranges and classes printed in BOLD and Cursive, are NOT available for 
installation in North America 

TM-SH sensors are restricted to Tamb = 60 °C and T process = -40°C to 60°C, see Table coils. 

Remote mount configuration (Junction box) 

Neck extension length 
(Between the TM, or TMU 
sensor neck, and the J-Box) 
Sensors have a standard neck 
> 60mm 

Process temp. (°C) 

-50 °C to .. 

Ambient temp. 

range (°C) 

-40 °C to .. 

Temperature 

class 

without 

40 

40 

T6 

without 

60 

60 

T5 

without 

100 

80 

T4 

100 mm 

115 

80 

T4 

100 mm 

175 

80 

T3 

100 mm 

180 

80 

T2 

200 mm 

220 

80 

T2 

CAUTION: The temperature ranges and classes printed in BOLD and Cursive, are NOT available for 
installation in North America. 

Содержание Heinrichs TM

Страница 1: ...AL FOR EXPLOSION PROOF FLOWMETERS EX Supplement MANUAL TM TMU TM SH March 2021 Coriolis mass flow sensors TM TMU TM SH Supplementary Operating Manual for explosion proof flow sensors Please read the i...

Страница 2: ...nsor transmitter 8 3 4 EC Type Examination Certificate for the sensors type TM TMU and TM SH 9 3 5 Electrical Parameters 10 3 5 1 Exciter circuit terminals 8 and 9 10 3 5 2 Sensor circuits terminals 1...

Страница 3: ...h 2021 or later These instructions are supplementary operating manual for non explosion proof Coriolis flow sensors If you do not have a copy of the latter instructions please request one from Heinric...

Страница 4: ...maintenance are to be performed by a technician trained to work with explosion proof devices or by a Heinrichs Messtechnik service technician Warning Any safety relevant maintenance or repairs in term...

Страница 5: ...itized using the terms listed below which are defined as follows in regard to the instructions herein and the advisories pertaining to the device itself Warning means that failure to take the prescrib...

Страница 6: ...no responsibility for the selection of unsuitably materials Caution To ensure the device performs correctly and safely it must be shipped stored set up mounted operated and maintained correctly 2 Ide...

Страница 7: ...rom Heinrichs Messtechnik during your order process Although the appearance of the standard and explosion proof transmitters are often identical and their rating plates contain similar information und...

Страница 8: ...rfaces pursuant to EN 1127 1 Explosive atmospheres Explosion prevention and protection paragraphs 5 2 and 6 4 2 3 3 3 Connection remote sensor transmitter A dedicated cable SLI2Y SP CY 5 x 2 x 0 5 mm...

Страница 9: ...ficate for the sensors type TM TMU and TM SH The sensors are certified with the following examination certificates CSA North America CSA18CA70171067X Class I Division 1 and 2 Group A B C D Class I Zon...

Страница 10: ...Li 4 38 mH 3 5 1 2 For exciter circuit type EC2 For connecting an intrinsically safe circuit with an Ex ia IIC type of protection linear output characteristic and the following maximum values Voltage...

Страница 11: ...without 40 45 T6 without 60 60 T5 60 mm 100 80 T4 160 mm 115 80 T4 160 mm 175 80 T3 160 mm 180 80 T2 260 mm 220 80 T2 CAUTION The temperature ranges and classes printed in BOLD and Cursive are NOT ava...

Страница 12: ...ables are to be installed in such a way that they are protected against mechanical damage and unduly high temperatures The external diameters of the connecting cables must be compatible with the thick...

Страница 13: ...t the equipment is only installed in areas which comply with the approved types of protection and environments e All connected electrical equipment must be suitable for its intended use f The operator...

Страница 14: ...t an ambient temperature of less than 20 C or greater 60 C suitable cables cable entries and conduit entries are to be used i The operational wall thickness of the stainless steel flow tubes may be 1...

Страница 15: ...nt by opening the enclosure and or using where necessary tools and test equipment a Maintenance or replacement work must be carried out by qualified personnel only i e personnel qualified according to...

Страница 16: ...es Such equipment shall only be repaired by expert personnel trained and authorized to do so Damaged threaded or flat joints which may impair the IP protection of the enclosure shall not be repaired o...

Страница 17: ...CAN CSA C22 2 No 60079 11 14 Explosive atmospheres Part 11 Equipment protection by intrinsic safety i ANSI ISA 60079 0 2020 Explosive Atmospheres Part 0 Equipment General Requirements ANSI ISA 60079...

Страница 18: ...Heinrichs Messtechnik GmbH Page 18 of 20 TM TMU TM SH Supplement Manual For Explosion Flow meters 7 EU Declaration of conformity...

Страница 19: ...Heinrichs Messtechnik GmbH TM TMU TM SH Supplement Manual For Explosion Flow meters Page 19 of 20...

Страница 20: ...Heinrichs Messtechnik GmbH Page 20 of 20 TM TMU TM SH Supplement Manual For Explosion Flow meters...

Отзывы: