19 Trouble shooting
Diagnosis
Possible reason
Corrective action
Indicator window
steamed
- Indicator cover damaged
- Indicator sealing damaged
- Cable gland not tightened
- high temperature differences
- aggressive ambient atmoshpere
- change indicator cover
- tighten cable gland
- indicator with pressure relief
- install rinsing connection for N2 gas
Indicator cover untight
- indicator sealing damaged
- change indicator cover
- fasten indicator cover screws
Indicator window icy
- ambient temperature too low
- install N23 rinsing connection
- change to high tem. distance housing
Instrument shows wrong
measuring values
- calibration does not meet process
conditions
- pointer de-adjusted
- review operating conditions, on deviations new
calibrations from manufactuer is recommended
- re-adjust pointer (RP on meas. scale)
No reaction on pointer
despite flow
- pointer is behind the stopper ( RP)
- float is sticked
(pollution)
- remove cover, re-adjust pointern
- remove instrument from process and clean float
and all inner parts carefully
- if sticking parts in the process use of magnet filter
recommended
Float is sticked
- pollution / residues in the process
- Measuring pipe damaged
- gas or liquid damping system damaged by
the process
- remove instrument from the process and clean
carefully
- resend the instrument to the manufacturer for
evaluation
Pointer indication is
pulsating
- Inlet pressure too low
- increase inlet pressure, instrument for recalibration
back to manufacturer (gases) or
- use of instrument with low pressure loss
- re-equip with double turbulent flow damping
- locate valves, gate valves etc. after the instrument
Electrical equipment e.g.
transducer, limit swithes
without function
- wrong wiring
- defective el. parts
- check wiring diagram acc. instruction manual
- change el. parts ( e.g. limit switches ) or sent the
instrument back to manufacturer
Un-known
Sent the instrument back to the manufacturer and
include a detailed failure description
20 Returning devices for repair and service
Note:
In accordance with the applicable German waste disposal legislation, the owner/client is responsible for the disposal of special
waste and hazardous materials. Consequently, all devices sent to us for repair must be free of any hazardous materials. This also
applies to possible hollow spaces and fissures in the devices. If repair is necessary, confirm the above-mentioned item in writing
(please use the form in the Appendix)
.
If hazardous materials remain in or on the device after it has been returned, Heinrichs Messtechnik shall be authorized to remove them
at the client’s expense without further inquiry.
21. Replacement parts
Group
Description
Info
Anzeigeteil
Indicator cover aluminum c/w window/gasket/screws
Scale, blank
Scale incl customer operating values
Indicator unit(B)-aluminum, c/w. mit blank-scale
Indicator unit(B)-aluminum, c/w. limit switches KEI1 (SJ3,5-N)
Indicator unit(B)-aluminum, c/w. limit switches KEI2 (SJ3,5-N)
Switching disc for KEI
El. transducer ES, 4-20mA, Hart®
Indicator unit(B)-aluminum, c/w. el. transmitter ES
Bearing unit c/w (B)-standard c/w.axle, bearings and magnets
Bearing unit (B) for indicator on distance
Pointer unit complete. (standard)
Pointer unti complete fol el. transducer ES
Base plate(B)-aluminum, c/w cover screw for cable entry(standard)
Base plate(B)-aluminum, c/w cover screw for cable entry(standard)
for indicator on distance
Cable gland M20x1,5, grau, Polyamid
Cable gland M20x1,5, blue, Polyamid
Flow tube
Float st.st. c/w guiding star
on request
Float st.st. c/w guiding star, and liquid or gas damping system
on request
Mounting
sets
Retrofit set for KEI1 (SJ3,5-N)
Retrofit set for KEI2 (SJ3,5-N)
NOTE
:
When ordering spare parts please always mention the serial number of the instrument.
If the float is being removed we recommend a new calibration of the instrument always. Without new calibration additional
measuring errors must be considered.
For further instruction on spare parts please contact the Heinrichs Messtechnik GmbH service department