ZED-Z
page 14
ZED-Z K11/0522
RELAY S1
Menu Item
Parameter / Function
Explanation / Values / Other
s1Funct.
Relay1 Function selection
Flow.: Monitoring of an adjustable flow value (s1SPoint).
TA Limit: Monitoring of an adjustable total volume (s1TA-Lim).
PA-Limit: Monitoring of an adjustable part volume (s1PA-Lim).
see
8.0 Relay functions
s1Char.
Relay1 Switch characteristic
Limit: Monitoring a reading (s1SPoint).
Window: Monitoring an adjustable measuring range
(s1SPoint...s1Fpunkt).
(only if
s1Funct.
is set to
Flow
)
s1Logic
Relay1 Switch logic
normal: Relay 1 activated when the limit value is exceeded.
invers: Relay 1 drops out when the limit value is exceeded.
s1SPoint
Relay1 Switchpoint
Reading is in the same units as the flow indicator.
(only if
s1Funct.
is set to
Flow
)
s1Hyste
Relay1 Hysteresis
Reading is in the same units as the flow indicator.
(only if
s1Funct.
is set to
Flow
)
s1FPoint
Relay1 Windowpoint
Reading is in the same units as the flow indicator.
(only if
s1Funct.
is set to
Flow
und
s1Char.
auf
Window
)
s1SDelay
Relay1 Switch delay
Delays the switching of the relay when the limit value is exceeded.
Range:_ 0,0...99,9 sec
(only if
s1Funct.
is set to
Flow
)
s1RDelay
Relay1 Reset delay
Delays the switching of the relay when the limit value is undershot.
Range:_ 0,0...99,9 sec
(only if
s1Funct.
is set to
Flow
)
s1PA-Lim
Relay1 Part volume limit value
Limit value for monitoring the part volume counter.
Reading is in the same units as the part volume unit (PA-Unit.) in
counter menu group.
(only if
s1Funct.
Is set to
GM Limit
)
s1TA-Lim
Relay1 Total volume limit value
Limit value for monitoring the total volume counter.
Reading is in the same units as the total volume unit (TA-Unit) in
counter menu group.
(only if
s1Funct.
Is set to
TA Limit
)
Содержание DF ZL Series
Страница 1: ...Operating Instructions for Counter Electronic Model ZED Z or DF ZLxxx or Model ExxR ...
Страница 10: ...ZED Z page 10 ZED Z K11 0522 7 4 General Settings ...
Страница 12: ...ZED Z page 12 ZED Z K11 0522 7 5 Flow analogue output and relay 1 ...
Страница 15: ...ZED Z ZED Z K11 0522 page 15 7 6 Relay 2 and volume counter ...