APS-Z/APK-Z/ARS-Z
APS-Z/APK-Z/ARS-Z K04/1122
page 7
Second step
Immerse the pH electrode in a second buffer solution (for example pH 4.0
or 10.0) and wait until the display value has stabilized. Then enter the pH
value of the second buffer solution.
The measuring instrument determines the zero point and slope of the
sensor.
Rinse off the electrode with deionized water and carefully dry it with a soft
cloth.
The calibration is complete.