2.  Note 

Please read these operating instructions before unpacking and putting the unit into 
operation. Follow the instructions precisely as described herein.  
The instruction manuals on our website www.kobold.com are always for currently 
manufactured version of our products. Due to technical changes, the instruction 
manuals available online may not always correspond to the product version you 
have  purchased.  If  you  need  an  instruction  manual  that  corresponds  to  the 
purchased  product  version,  you  can  request  it  from  us  free  of  charge  by  email 
([email protected]) in PDF format, specifying the relevant invoice number and 
serial number. If you wish, the operating instructions can also be sent to you by 
post in paper form against an applicable postage fee.

 

 
The devices are only to be used, maintained and serviced by persons familiar with 
these operating instructions and in accordance with local regulations applying to 
Health & Safety and prevention of accidents.  
 
When  used  in  machines,  the  measuring  unit  should  be  used  only  when  the 
machines fulfil the EC-machine guidelines. 
 
If you encounter difficulties during startup, please do not interfere with the electrode 
in any way that is impermissible. By doing so, you could endanger your rights under 
the instrument warranty. Please contact your supplier. 

  Each electrode is a quality product and is individually tested and packaged. 

  Storage  should  be  in  a  dry  room  at  -5  to  +30°C.  Because  pH  and  redox 

electrodes have limited suitability for storage, we do not recommend storing 
them for longer than 6 months. 

  All electrodes are provided with a rinsing cap or container for transport and 

storage to prevent the sensor elements from drying out. 

  The electrode should be kept in a 3mol/l solution of KCl. 

  If the electrode has been stored dry for an extended time, however, it must 

be conditioned before being used for measurements. For this purpose, the 
electrode is immersed in a 3mol/l solution of KCl for about 24 hours. 

  The  pH-sensitive  membrane  glass  must  be  handled  carefully  (no  skin 

contact, protect against damage, etc.)! 

  Make certain that electrical connections and cables are kept clean and dry. 

 
If you send the electrode back, we need a brief description of the fault: 

o

  Electrode cannot be calibrated 

o

  Display is not stable 

o

  Other fault: 

 
 
 
 
 
 
 

Содержание APK-Z

Страница 1: ...Operating Instructions for pH Redox Electrodes Combination Electrodes Model APS Z APK Z ARS Z...

Страница 2: ...plemented in later versions The described products can be improved and changed at any time without prior notice Copyright All rights reserved 1 Contents 1 Contents 2 2 Note 3 3 Instrument Inspection 4...

Страница 3: ...used only when the machines fulfil the EC machine guidelines If you encounter difficulties during startup please do not interfere with the electrode in any way that is impermissible By doing so you c...

Страница 4: ...inspection of the delivery packaging In case of damage please inform your parcel service forwarding agent immediately since they are responsible for damages during transit Scope of delivery The stand...

Страница 5: ...tallation Electrodes must be installed vertically The maximum angle to the vertical is 80 The internal buffer must cover the inside surface of the membrane glass Air bubbles in the membrane chamber mu...

Страница 6: ...elected to bracket the later measurement range Rinse off the electrode with water between measurements First Step Connect the pH redox electrode with the transmitter and immerse it in a buffer or test...

Страница 7: ...for example pH 4 0 or 10 0 and wait until the display value has stabilized Then enter the pH value of the second buffer solution The measuring instrument determines the zero point and slope of the se...

Страница 8: ...ng If careful dabbing with a soft moist cleaning tissue proves unsuccessful various chemical cleaning methods can be used depending on the type of impurity Type of accumulation Cleaning agent Lime and...

Страница 9: ...4 days with monthly calibration Electrodes are naturally consumed by loss of salt from the reference system etc A drift in measured values after some time is therefore normal measuring behavior If the...

Страница 10: ...ions of usage Depending on the application the time until an electrode must be replaced could vary from a few days to several years Because of this it is not possible to offer a credible guaranty for...

Страница 11: ...g 13 5 1 general purpose glass 2 high temperature glass 5 135 C Order Details Connection cable Model Cable length Cable type APK Z 5 5 m cable length 1 10 m cable length K Standard coaxial cable R coa...

Страница 12: ...d international disposal regulations and environmental regulations Batteries Batteries containing pollutants are marked with a sign consisting of a crossed out garbage can and the chemical symbol Cd H...

Отзывы: