KMD-16
TRMS Multimeter
106
www.grupotemper.com
7.2 Replacingthefuses
WARNING:
To avoid electric shock, disconnect the test leads from any source of voltage before removing the fuse cover.
1. Disconnect the test leads from the meter.
2. Remove the battery cover.
3. Gently remove the old fuse and install the new fuse into the holder.
4. Always use a fuse of the proper size and value (0.5A/600V fast blow for the 600mA range, 10A/600V fast blow for the
20A range.
5. Replace and secure the rear cover, battery and battery cover.
WARNING:
To avoid electric shock, do not operate your meter until the fuse cover is in place and fastened securely.
8. Specifications
Function
Range
Resolution
Accuracy
DC Voltage
400 mV
0,1 mV
±(1.0%r 5digits)
4 V
0,001 V
±(1.0%r 3digits)
40 V
0,01 V
400 V
0,1 V
600 V
1 V
AC Voltage
50Hz to 60Hz
4 V
0,001 V
±(1.2%r 3digits)
40 V
0,01 V
400 V
0,1 V
600 V
1 V
All AC voltage ranges are specified from 5% of range to 100% of range
DC Current
400 μA
0,1 μA
±(1.0%r 3digits)
4000 μA
1 μA