91
Display
Action
Remark
The device returns to
measuring mode.
From the blinking
CAL mode indicator
you see that product
calibration has not
been terminated.
Product calibration
step 2
Display (3 sec)
Now the device is in
HOLD mode.
The stored value is
displayed (blinking) and
can be overwritten with
the measured sample
value.
Press enter to proceed.
Display of new
asymmetry potential
(based on 25°C).
Sensoface is active.
To end calibration:
Select MEAS, then
enter
Repeat calibration:
Select REPEAT, then
enter
Calibration is
terminated.
After end of calibration, the outputs remain in
HOLD mode for a short time.
Product Calibration (pH)