72
Configuration
Controller
(For description, see Controller Functions)
Process variable, controller type, setpoint.
1 Press any arrow key.
2 Select CONF using
keys, press enter.
3 Select REL1/REL2 menu using
keys,
press enter, then select CONTROLLER.
4 All items of this menu group are indicated by
the “CTR:” code.
Press enter to select menu,
edit using arrow keys (see next page).
Confirm (and proceed) using enter.
5 End: Press meas key until the [meas] mode
indicator is displayed.
Use of relays
Select process variable
Controller type
Pulse length
Pulse frequency
Setpoint
Neutral zone
Controller: P action
Controller: I action
Controller: D action
Behavior during HOLD
enter
4
MEMO SENS
1
MEMO SENS
MEMO SENS
2
3
enter
5
enter
4
meas
MEMO SENS
enter