Knick SE 554X AMSN Series Скачать руководство пользователя страница 3

SE 554X/*-AMSN

SE 554X/*-NMSN

SE 554X/*-NVPN

Manual

Instructions for Use for the Series SE 554 pH/ORP Sensors

1

Sensor connectors 

a - MS 

b - VP

2

19 mm A/F, serial number

3

PG 13.5 thread

4

PVDF compression ring 

5

EPDM-FDA O-ring (11.5 x 2.6 mm)

6

Rating plate

7

Junction 

8

pH glass

9

Platinum (ORP sensors only)

!

 WARNING – Failure to observe this warn-

ing may result in serious injury.

The warning symbol on the rating plate 

means:

Read these instructions for use, observe the Spec-

ifications, and follow the Safety Instructions. 

1  Safety Instructions

1.1 All Applications – All Sensor Variants

Hazards due to pressure, temperature, aggres-

sive media or explosive atmosphere are possi-

ble, depending on the location of use. There-

fore, the installation, operation, and servicing of 

the sensor shall only be carried out by suitably 

trained personnel authorized by the operating 

company. 

1.2 Hazardous Areas – All Sensor Variants

Observe all applicable local codes and stan-

dards for the installation of electrical equipment 

in hazardous locations. For orientation, please 

refer to IEC 60079-14, EU directives 2014/34/EU 

and 1999/92/EC (ATEX), NFPA 70 (NEC), ANSI/

ISA-RP12.06.01. The electrical and thermal pa-

rameters of the sensors must be adhered to.

1.3 Hazardous Areas – Sensors with Memosens 

Connector

Memosens Ex sensors are marked by an orange- 

red ring. Combined with a model CA/MS-***X** 

or model CA/MS-***X**-L measuring cable or 

a certified measuring cable which is identical 

in hardware and function, the sensor may be 

connected to a suitable measuring device, as 

described in the Certificates BVS 15 ATEX E141 X 

and IECEx BVS 15.0114X.

1.4 Hazardous Areas – Sensors with VP 

 Connector

The sensors shall only be operated with an 

approved device in floating, intrinsically safe 

circuits. 

2  Intended Use

The sensor is used for continuous measurement of pH 

and optionally of ORP (Model AMSN) in liquid media. 

The SE 554 is a low-maintenance (polymer electrolyte) 

and pressure-resistant sensor. It has a temperature 

detector for automatic temperature compensation. 

The sensor is designed for applications in industrial 

processes: 

•  Heavily polluted media

•  Media containing sulfides

•  Emulsions and suspensions

•  Alkaline media

•  Media containing proteins

•  Processes with large pressure variations

•  Acidic media

•  Media containing solvents

The sensor is non-sterilizable / non-autoclavable and 

unsuitable for measurements in media containing 
chlorine or bromine.

3  Installation and Commissioning

 •

On unpacking, check the sensor for mechanical 

damage. Report any damage to your Knick service 

team.

 •

Remove the watering cap and briefly rinse the sen-

sor with pure water. After rinsing, the sensor should 

only be dabbed dry with a tissue. Do not rub the 

pH-sensitive glass, since this can lead to electrostat-

ic charging and sluggish response times.

 •

Check the space behind the pH-sensitive glass for 

the presence of any air bubbles and remove them 

by gently swinging the sensor in a vertical plane.

 •

Install the sensor in the fitting as described in the 

user manual of the respective fitting.

 •

Connect sensor and cable.

4  Operation

4.1 Calibrating the Sensor

2-point pH calibration is recommended for the SE 554 

sensor: First remove the watering cap. Then dip the 

sensor successively into two different buffer solutions 

with given pH values (e.g., CaliMat pH 7.00 and pH 

4.00) and calibrate the sensor to these buffer values. 

ORP calibration is performed as a 1-point calibration 

(delta value) using an ORP calibration solution. Please 

refer to the user manual of the pH transmitter for fur-

ther details. 

4.2 Temperature Detector

The integrated temperature detector is intended for 

automatic compensation of the pH signal and not for 

any high-precision and safe temperature indication 

or control of the process temperature. 

5  Maintenance and Cleaning

Carefully rinse the sensor tip and junction with pure 

water after each operating cycle. Under no circum-

stances must measuring solution be allowed to dry 

on these parts! 

When the sensor is not in operation, store it with sen-

sor tip and junction well submerged in electrolyte 

(3 mol/l KCl). If a sensor is stored dry for a few days 

by mistake, let it soak in electrolyte for several hours 
before use.

6  Specifications

Model Code

The markings on each sensor or on the packaging 

label include the following information:

SE 554X/ 1 - AMSN 

Model designation 

  

Sensor connector 

   

MS: Memosens®

   

VP: VarioPin

  

Pt electrode (ORP)

   

A: with

   

N: without

  

Length 

   

1: 120 mm

   

2: 225 mm

   

3: 325 mm

   

4: 425 mm

  

Ex approval

  

  X: Yes

Further Data
pH/ORP range

0 ... 14 / ±1500 mV

Temperature

0 … 130 °C

Pressure, relative

0 ... 10 bar

Junction

2 x hole

Electrolyte

Solid polymer

Reference system

Ag/AgCl

pH sensor material

Alpha glass

medium impedance,  

universal glass,  

fluoride resistant

ORP sensor material Platinum
Body material

Glass

Mounting

PG 13.5

Temp detector

MS connector: NTC 30 kΩ

VP connector: Pt1000

7  Disposal

Observe the applicable local or national regulations 

for disposal.

pH

1b

8

Ø 12 mm

MEMO

SENS

5

4

3

6

120 / 225 / 325 / 425 mm

2

9

ORP

7

TA-SE554X-AMSN-KNX02  20180529

094597

Knick Elektronische Messgeräte GmbH 

& Co. KG 

Beuckestraße 22

14163 Berlin

Germany

Phone:   +49 30 80191-0

Fax:  

+49 30 80191-200

Email:   [email protected]

Internet: www.knick-international.com

1a

Содержание SE 554X AMSN Series

Страница 1: ...Ein Abreiben des pH sensitiven Glases kann die An sprechzeit durch elektrostatische Aufladung stark erhöhen Entfernen Sie gegebenenfalls Luftblasen im Raum hinterdempH sensitivenGlasdurchleichteSchleu derbewegungen in senkrechter Ebene Setzen Sie den Sensor in die Armatur ein wie in der Anleitung der Armatur beschrieben Verbinden Sie Sensor und Kabel miteinander 4 Betrieb 4 1 Kalibrierung des Sens...

Страница 2: ...stung der extern angeschlossenen Stromkreise und beträgt 427 K W in Luft Die maximal zulässige Umgebungstemperatur beträgt 40 C Temperaturklasse Maximale Eingangsleistung Pi Maximal zulässige Prozesstemperatur T6 50 mW 55 C T4 110 mW 80 C T3 200 mW 110 mW 100 C 135 C Besondere Bedingung ohne Für Sensoren mit Memosens Steckkopf Bescheinigungsnummer Kennzeichnung BVS 16 ATEX E 037 X IECEx BVS 16 003...

Страница 3: ...t any damage to your Knick service team Remove the watering cap and briefly rinse the sen sor with pure water After rinsing the sensor should only be dabbed dry with a tissue Do not rub the pH sensitiveglass sincethiscanleadtoelectrostat ic charging and sluggish response times Check the space behind the pH sensitive glass for the presence of any air bubbles and remove them by gently swinging the s...

Страница 4: ...y connected circuits and is 427 K W in air The maximum permissible ambient temperature is 40 C Temperature class Maximum input power Pi Maximum permissible process temperature T6 50 mW 55 C T4 110 mW 80 C T3 200 mW 110 mW 100 C 135 C For Sensors without Temperature Detector S8 Connector The maximum heating to be considered with regard to the temperature class is negligibly low The maximum permissi...

Отзывы: