background image

Manual

5  Maintenance, Cleaning, Storage

Dirt deposits on the junction and the glass membrane 

can lead to both a sluggish response and to measure-

ment errors. 

Use suitable agents to clean the sensor:

Water-soluble  

impurities

Water

Grease, oil 

Warm water with household 

detergent,alcohol 

Lime

Acetic acid or  

0.1 mol/l hydrochloric acid

Protein

Pepsin/hydrochloric acid 

solution

Note:

 Never wipe the sensor with a dry cloth as it 

may then be unusable for some time due to static 

charging.

When you do not use the sensor, store it with the wa-

tering cap attached. It should be filled with 3-molar 

KCl solution (ZU 0960).

6  Specifications

Measuring range

pH 2 ... 13

Temperature

-5 ... 80 °C

Junction

Open junction

Electrolyte

Polymer

Sensor connector

Memosens (MS)

Connector material

PPS/PVC

Body material

Noryl (PPE)

Temperature detector

NTC 30 kΩ

7  Package Contents

 •

Sensor

 •

Instructions for Use

8  Disposal

Observe the applicable local or national regulations 

for disposal.

SE 104-MS

Instructions for Use for the SE 104-MS pH Puncture Sensor

1

Sensor connector (Memosens®)

2

Gasket for Portavo quiver

3

Nameplate

4

Body

5

Junction

6

pH glass

Read these instructions for use, observe the Spe-

cifications, and follow the Safety Instructions. 

The SE 104-MS puncture sensor is equipped with 

Memosens technology. The benefits are a safe func-

tion due to perfect galvanic isolation, digital data 

transmission and an integrated sensor diagnostics, 

among others. The connection is insensitive to dirt 

and humidity. The calibration data, operating time, 

sensor designation, and serial number are stored in 

the sensor head.

1  Safety Instructions

Make sure that the sensor is without visible mechani-

cal damage. Use the sensor for the intended purpose 

only.

2  Intended Use

The sensor is designed for simultaneous measure-

ment of pH and temperature in industrial processes, 

particularly in semi-solid materials. Observe the per-

missible operating conditions.

3  Installation and Commissioning

 •

On unpacking, check the sensor for mechanical 

damage. Report any damage to your Knick service 

team.

 •

Connect the sensor with the Memosens cable and 

then connect the Memosens cable to the measur-

ing device.

4  Operation

4.1 Calibrating the Sensor

Calibrate the sensor regularly – ideally prior to each 

measurement. The frequency of calibrations is highly 

dependent on the required measuring accuracy. Only 

use unspoiled buffer solutions (CaliMat) within their 

expiry date.

Note: 

Calibration buffers are consumables. Use each 

buffer solution only once.

Two-point calibration is recommended for the SE 

104-MS sensor. Please refer to the user manual of the 

meter for further details. 

4.2 Measure 

In addition to the pH value, the sensor also measures 

the temperature. The temperature measurement is 

used for temperature compensation of the sensor, 

which is a prerequisite for correct pH measurement. 

To document the measurement results, you should 

specify the measured pH value together with the 

measuring temperature.

CAUTION: 

Prevent the sensor from hitting hard par-

ticles in the measured medium as this can destroy 

the sensor.

4

1

3

2

MEMO

SENS

5

6

Knick Elektronische Messgeräte  

GmbH & Co. KG

 

Beuckestraße 22

14163 Berlin

Germany

Phone:  +49 30 80191-0

Fax:  

+49 30 80191-200

[email protected]

www.knick-international.com

Отзывы: