72
Appendix:
Minimum Spans for Current Outputs
The OXY 3400(X)-067 module is a measuring module. It does not provide cur-
rent outputs. Current outputs are provided by the BASE module (basic device)
or by communication modules (e.g. OUT, PID). The corresponding parameters
must be set there.
The minimum current span shall prevent that the resolution limit of the
measurement technology (± 1 count) is seen in the current.
OXY 3400(X)-067 Module
%Air
0.1
%O
2
0.1
°C
10.0
mbar
20.0
(barometric pressure)
nA
10 % min. 1.00 nA
mg/l
2 µg/l
ppm
2 ppb
mbar
1 mbar
(partial pressure)
Vol%
0.01
ppm
100
°F
10.0
Calculation Block OXY/OXY
Diff %Air
0.1
Diff-%O
2
0.1
Diff mg/l
10 % min. 2.0 µg/l
Diff ppm
10 % min. 2.0 ppb
Diff °C
10.0
Diff Vol%
0.01
Diff ppm
100