Diagnostics
Sensor wear monitor
23.0 °C
Return
Sensor operating time
Autoclaving cycles
98%Air
SIP cycles
CIP cycles
Membrane wear
312 d
1 of 2
0 of 3
1 of 5
Inner body wear
Sensor Wear Monitor
The Diagnostics menu provides information on
the current wear of inner body and membrane
body.
Generally, the membrane body must be
replaced more often than the inner body:
The “Membrane wear” bargraph shows how
much the membrane is worn out.
The inner body wear is not reset when a mem-
brane body has been replaced - here, the load
on the inner body is added up and represented
as total load on the inner body.
Sensor network diagram OXY
Return
23.0 °C
98%Air
2 - Zero point
1 - Slope
4 - Not in use
3 - Sensocheck
5 - Response time
7 - Sensor wear
6 - Cal timer
Sensor Diagram
• Slope
• Zero
• Sensocheck
• (Not in use)
• Response time
• Cal timer
• Sensor wear
The measured values are continuously moni-
tored during the measurement process. The
sensor network diagram provides at-a-glance
information about critical parameters. If a toler-
ance limit has been exceeded, the respective
parameter is flashing.
Values in gray: Monitoring switched off.
Critical range –
“inner circle”
Value out of tolerance
The tolerance can be
modified as required!
“Outer circle”
Value within tolerance