16
Select AI-TB configuration:
Assign each process variable to one
of the 4 Analog Input Blocks per TB.
(See Page 19)
NOTICE!
The assignments made
here also apply to the FF configura-
tion for the connection of Analog
Input Blocks to the channels of the
AI Transducer Blocks! Be sure to make
identical settings!
AI TB1 configuration (Administrator)
Return
7.00 pH
25.6 °C
Analog Input AI 2
pH
°C
mV
°F
Analog Input AI 1
Measuring module
Analog Input AI 4
Analog Input AI 3
Module PH 3400-032
AI-TB Configuration on the Device
Assigning process variables to Analog Input Blocks on the device
Menu Display
Assigning process variables to
Analog Input Blocks
Bus Activation
On:
Access control by DCS
(Control system programming
required by customer!)
Off:
Device access via passcode or
signature (Audit Trail) as usual
Bus activation (Administrator)
Return
7.00 pH
25.6 °C
Bus activation
Off
On
Note:
If access control has inadvertently
been activated, the device must be
separated from the FF bus to make
the menu accessible again.
Menu Display
Bus activation
N
ot
e:
T
he displa
y ma
y v
ar
y depending on the devic
e v
ersion.