40
Error Codes and Device Messages
Error Messages
The following error messages can be shown in the display.
Message
Cause
Remedy
blinks
Battery empty
Replace batteries
ERROR 1
Value out of range
Check whether the measurement
conditions correspond to the
adjusted measuring range.
ERROR 3
Temperature value out of
range
ERROR 6
Cell constant too high/low
Enter nominal cell constant or
calibrate the sensor using a known
solution.
ERROR 11
Measured value unstable
Drift too high
Leave the sensor in the liquid until
the temperature is stable. If this does
not help, replace the sensor.
ERROR 14
Time and date invalid
Set time and date
ERROR 18
Configuration invalid
Restart, reset to factory settings
(Setup: DEFAULT YES), configure and
calibrate.
If this does not help, send in the
device for repair.
ERROR 19
Factory settings error
Device defective, send it in.
ERROR 21
Sensor error (Memosens)
or
Sensor verification message
Connect an operational Memosens
sensor.
With sensor verification activated in
Paraly SW 112, this error message
indicates that an unassigned sensor
was connected.