21
Calibrating
CAL SOP Calibration
(Option, must have been configured in Paraly SW 112 software)
Measurement
cal
CAL
CAL SOP blinks
Use
st
to select the required calibration
method (CAL SOP).
cal
Perform the selected calibration
(see Paraly SW 112 software).
Option 001 SOP (Standard Operating Procedure) meets specific requirements of the
pharmaceutical and biotechnological industries.
In the Paraly SW 112 software, you specify which buffers are to be used in which
sequence. You can combine buffer solutions from different buffer sets.
Please note that the minimum distance allowed between two buffer solutions is
Δ 2 pH.
SOP calibration allows you to:
• use up to 3 buffers for adjustment
• use a 4th point for verification (“verification buffer”)
• specify a maximum deviation from the verification buffer
• use buffers from different buffer sets, including a “user buffer”.