17
Calibrating
0.1 / 0.01 MOL KCL Calibration
(Automatic calibration with KCl solution)
The calibration method is selected in the configuration menu.
Measurement
CAL
0.1/0.01 MOL KCL
PRESS CAL
Dip sensor in KCl solution.
The meter automatically compensates for the
temperature deviation!
Measured value
Temperature
Conductivity KCl
Hourglass blinks
Calibration is performed. Automatic return to measuring mode.
NOTICE!
• Make sure that the values of the calibration solutions used correspond exactly
to those specified in this manual.
If not, the resulting cell constant will be incorrect.
• When calibrating in a liquid, make sure that the sensor, the separate tempera-
ture probe (if present) and the calibration solution have the same temperature.
Only this ensures that the cell constant is determined correctly.