Connecting the outlet hose
Turn the outflow nipple so that the outlet hose
points outwards (see figure on the left). Before
tightening the coupling nut, correctly align and
insert the outflow nipple!
Outlet hose
Coupling nut
Coding recesses
Coding bars
Outflow nipple
Information on Installation
Installing the sensor cable
(pH, oxygen, and conductivity cables)
Attach the connecting cable
(A)
to the sensor and
connect to the fixing bracket
(B)
in a loop as shown in
the illustration.
Using clamp
(C)
, attach the cable lightly so that the
cable is secured and not constricted. The rotational
movement of the Ceramat would otherwise damage
the cable.
A
B
C
F
NOTICE
Check positioning of coding recesses and coding
bars to lock the connection.
NOTICE
The cable loop must be long enough so that the
cable does not impede the stroke movement of
the fitting.
Connect equipotential bonding cable to terminal
(F)
(if required).